Figure 15. Clean the Sides of
the SEM Scanner 200
Figure 16. Clean the Top
of the SEM Scanner 200
Figure 17. Clean the
Bottom of the SEM
Scanner 200
2.
Use the second wipe to completely clean the scanner again for 30 seconds.
3.
Use the third wipe to perform another 30 second overall final wiping of the device. The
device surface should be fully coated with the wipe solution after the cleaning
4.
Allow the device to sit for at least 2 minutes to properly disinfect the device before
returning it to storage or using on another patient.
5.
The charging mat is typically used in a clean office environment with clean and disinfected
scanners. The charging mat should be cleaned only as needed to maintain a good
appearance and proper functionality. More extensive cleaning may be required if the
system is accidentally soiled or contaminated. Follow the cleaning and disinfection
recommendations above.