Accuracy
Other
DC current and voltage ranges:
AC voltage ranges:
at reference conditions
DIN 43 780 Class 1.5 for DC (relative errors on the scale length, 77 mm)
and 1.5
for class sizes,
changing resistance reference conditions Ambient tempe20 ° C
Mounting position horizontal frequency 40 ... 500 Hz
Effect sizes according to DIN 43 780 MA 4 The S has a branch rectification and
is calibrated in rms values for sinusoidal AC 50 Hz
Nominal parameters and operating ranges
Temperature Temperature Error
Working temperature range frequency
0. . . +20. . . +40 ° C.
typical
± 0.7
/ O/10 Κ in the range 0 ... +40 ° C
-25. . . +40 ° C
15 .. 40 .. 500.. 10 000 Hz
Range 750 V:
15. . . 40 ... 500. .3 000 Hz
The frequency is influenced by grounding the
T
erminal "J_"
In the given reference ranges
corresponds to the basic error of class 1.5, in the spheres of influence,
the typical additional error ± 1.5%
<, 0.7% of full scale at an interference field strength of 0.5 mT <0.5% full scale
for open input terminals and ground terminal J_
(interference field strength 40 kV / m, frequency 50 Hz)
error <± 1.5% of full scale when the AC
component is not greater than the selected upper range limit AC current ranges is:
Error <± 2.5% of full scale when the DC component is not greater than the selected
upper range limit
error <± 2.5% of full scale when the DC component is not
greater than 50% of the selected measuring range value
Maximum allowable stresses and Thresholds of the guards must not be exceeded!
When the measured variable rectangular display is 11% too large,
triangular-shaped in measured quantity by 4% too small.
%
%