Extras
44
Need Help ? Bayer Diabetes Care: 1-800-348-8100
(available 24 hours a day, 7 days a week)
Extras
45
www.bayerdiabetes.com/us
Technical Information
Specifi
cations
Specifi cations:
Test Sample:
Whole blood
Test Result:
Referenced to plasma/serum glucose
Sample Volume:
0.6 µL
Measuring Range:
20- 600 mg/dL
Measuring Time:
5 seconds
Memory Feature:
Stores most recent 480 test
results
Radio Frequency:
916.5 MHz
Battery Type:
Two 3-volt lithium batteries
(DL2032 or CR2032)
Battery Life:
Send
option off:
About 1000 tests (1 yr.
average use).
Send
option on:
About 244 tests (2 months at
4 transmitted results per day).
Meter Life:
5 years
Operating Temperature Range:
41°F–113°F
Humidity:
10–93% RH
Altitude:
Up to 11,000 feet does not signifi cantly
affect results.
Dimensions:
77 mm (H) x 57 mm (W) x 23 mm (T)
Weight:
1.85 ounces
Alarm:
“Beeps” whenever a test strip is inserted into
or removed from the meter, when the test strip is fi lled
with blood or control solution, and when a test result
appears in the display. Two beeps will sound for an
error. Twenty beeps will sound when a programmed
alarm goes off.
To assess intermediate measurement precision, the
performance of ten
C
ONTOUR
®
L
INK
meters was evaluated
with three lots of sensors. The evaluation was performed
by three operators who utilized three liquid control glucose
levels under normal conditions of use over the course of 10
days. In total, three hundred sensors were used for each
glucose level. The Grand Mean, pooled standard deviation
and pooled %CV for each glucose level were calculated
and are presented here.
Glucose Level
(mg/dL)
Glucose" Level
(mg/dL)
Pooled SD
Pooled CV (%)
Low (30-50)
39.9
0.8
1.9
Normal (96-144)
123.6
2.1
1.7
High (280-420)
360.7
6.6
1.8