41
What You See
What It Means
What You Do
Temperature is either too
hot or too cold.
Allow the meter to reach
room temperature.
Test strip removed during
the test, or release button
pressed during the test.
Take care to not press
release button or remove
test strip during test.
Release used test strip and
repeat test (see pg. 8).
The meter is open when
the meter handle is pulled.
Close meter and push
handle in.
Meter cannot perform
automatic calibration,
and you must enter
the Program Number
manually.
Check test strip disc
or carton for Program
Number (P#). Use the
or
to select the
correct number and press
to accept.
Battery life has ended.
Replace the battery.
Low Battery icon flashes if
battery power is low.
Replace battery within 20
readings or within one
week.