10. CALIBRATION (ONLY AUTHORIZED PERSONAL)
The calibration can be only done by the authorized technical personnel, please, if you wish do a calibration
contact with your authorized calibration organism who is authorized to seal and to unseal the scale.
In order to enter the calibration parameter it is necessary to short the K2 jumper K2 of the main keyboard.
Turn the power off.
Turn the power back on, during the counting from 9 to 0 press the
key.
The display will show
Pn
(ask password), press
to enter.
The display wil show the first function,
PO CHk
. Press
until display show
P2 CAL
, press
to enter,
press
to select
CAL
, press
key to enter calibrate. The display will show
UnLd
.
Remove any weight from the platform. After stable indicator on, press the
key.
Then the display will show the last calibration weight used. If this is correct you can continue by pressing
the
key. If it is not correct use the
keys to change the calibration weight value. When it
is correct press the
key.
Then display will show
LoAd
. Place the calibration weight on the scale. After stable press the
key.
If the calibration is acceptable the display will return to normal. If an error message is shown try calibration
again as a disturbance may have prevented a successful calibration.
If the problem persists then contact your dealer.
After calibration, itshould be checked to verify the calibration and linearity is correct. If necessary repeat
calibration, especially be certain the scale is stable before accepting any weight.
11. ERROR CODE
ERROR
CODES
DESCRIPTION
RESOLUTION
- - - - -
Over range
Remove weight from the indicator.
If the problem persist contact your dealer for assistance.
Err4
Zero Setting Error
The indicator was outside the normal zero setting range either when it was
turned on or when the
key was pressed.
Remove weight from the scale and try again.
Use the
key to set the display to zero value.
If the problem persist contact your dealer for assitance.
Err6
A/D out of range
The values from the A/D converter are outside the normal range. Remove
weight from the scale if overloaded, make sure the pan is attached.
Indicates the load cell or the electronics may be faulty.
If the problem persist contact your dealer.
12. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the de
-
livery date.
During this period, GIROPÈS SL will take charge of repairing the scale.
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation
contrary to the recommendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.