EN
USER’S MANUAL
BR20 | BR30
26
In reading data lock mode, if you add/remove big weight, display will still update and lock new reading
data.
You can press
and
together to enter/escape animal weighing mode.
6.7. KEYBOARD LOCK
You can use keyboard lock when this function enable (see detail setting process in section 7), after keyboard
haven’t use for 10 minutes, keyboard will be lock, after enter lock status, if you press any key, display will
show
k-LCk
.
If you want to escape lock mode and return work mode, hold
,
,
key 2 seconds, display will
show
ULCk
, return normal mode.
6.8. SET BACKLIGHT
Hold
key 3 second, display will show
SEtbL
, press
key to enter backlight setting, press
key to
change backlight mode (
bL on
: always on,
bL AU
: auto backlight, backlight will on when press any key or
add load on scale,
bLoFF
: backlight always off).
Press
key to sure, press
key to scape.
6.9. SET AUTO POWER OFF
Hold
key 3 second, display will show
SEtbL
, press
key to enter.
Press
key, display will show
SEtoFF
, press
key to enter auto power off setting, pr
ess
key to
change auto power off tim
e
oF on
: always on,
oF 5
: auto power off after standby 5 minuts,
oF 15
: auto
power off after standby 15 minuts. Press
key to sure, press
key to scape.
7. PARAMETERS
The indicator has 3 parameters that can be programmed by the user and a parameter to do the calibra
-
tion.
The calibration can be only done by the authorized technical personnel, please, if you wish do a calibration
contact with your authorized calibration organism who is authorized to seal and to unseal the scale.
In order to enter the calibration parameter it is necessary to short the K2 jumper K2 of the main keyboard.
To set parameters press the
key during selfchecing, display whill show
Pn
(ask password), press
,
,
to enter.
The display will show the first function,
PO CHk
.
Pressing the
key will cycle through the other functions.
Pressing
will allow you to set the function. It may be necessary to either use
or set a value using
the
key to move the active digit and then using the
key to increment a digit, followed by the
key to enter the value. Use the
key to leave a parameter unchanged.
For example when the display shows
PO CHk
press the
key to begin.
This display will show
SEt Lo
, press the
key to set the low limit, or press the
to skip to the next
parameter,
SEt H i
for setting the high l
imit.
After pressing the
key to set a limit, use the
and
keys to change the flashing digit, then use
the
key to increment the flashing digit. Continue to the next digit and set it as needed.
When all digits have been set press the
key to store the value. The display will go back to the parameter
just st, i.e.
SEt Lo
. Advanced to another prameter if needed or press the
key to return to weighing.
Parameter setting table:
Содержание BR20