USER’S MANUAL
ARN
22
4.3 ERROR MESSAGES
DISPLAY
-O.L-
The weighing system has been overloaded. Remove weight.
Err06
Signal from the sensor on AD converter is too high. Check sensor connection..
Err07
No signal from the sensor on AD converter. Check sensor connection..
Err12
Span calibration data lost. Recalibrate the system.
Err98
Calibration point must be higher than previous one and higher than value of 25.
Or enterered value calibration point value is equal to 0000..
Symbol is blinking:
The battery voltage level is running low. The battery has to be charged.
Flahing M:
memory is full ( >99.999 kg cannot be shown by using 5 digits), no weights
can be accepted te be added. Erase memory.
4. 4 ON / OFF SWITCH
SWITCHING ON THE INDICATOR
Start up the indicator. Shortly press the Key
SWITCHING OFF THE INDICATOR
Switch off the indicator by keeping both keys
and
simultaneously pressed for 2 seconds.