– 14 –
Operating Hints
• To interrupt the water delivery, press STOP/0.
• To continue dispensing, press START/RESUME.
• To cancel dispensing, press SET AMOUNT.
• The amount of water dispensed is displayed in the WATER AMOUNT DELIVERED window. When
the delivery is complete, the display will blink and indicate 0 POUNDS 0 OUNCES.
• If the set temperature is not reached within the preset time allowed (see Flush Time), the meter will
fl
ash and beep until a new temperature is set, the amount is changed or the START/RESUME button
is pressed. If the START/RESUME button is pressed, the meter will deliver the set amount.
• To get better results, always discard the
fi
rst water sample when the new temperature is entered
or the unit has not been operated longer than 15 minutes.
USING THE TEMPERATURE PROBE
1. Verify that the hand held temperature probe is plugged into the end of the
fl
exible cable attached to
the bottom panel of the meter.
2. Place the end of the probe where the temperature needs to be measured. The HAND HELD PROBE
TEMPERATURE window will display the temperature reading.
NOTE:
The probe can be used to measure the temperature of the room, dough or
fl
our in the range of
32°F (0°C) to 125°F (52°C). The probe must be in place for 30 seconds to get an accurate temperature
reading.
Locking and Unlocking the Change Units
The CHANGE UNITS key can be locked to prevent it from being changed. To access Locking and Unlocking
the Change Units parameter, follow the steps below:
1.
Press
SET
TEMP.
2. Enter 32 using the numeric keypad.
3.
Press
SET
AMOUNT.
4. Enter 94404 using the numeric keypad. The WATER AMOUNT SETTING window displays
944 lbs 04 oz.
5. Press START/RESUME. The WATER TEMPERATURE SETTING window displays
0
(Units is not
locked) or
1
(Units is locked).
6. Enter desired value on the numeric keypad. If no change is desired, proceed to the next step.
7. Press START/RESUME to save the setting.