3. Warnings and Cautions
3-4
HomeChoice
APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
The Minimum Drain Volume % is programmed too low. This can cause
your Drain cycles to end early.
Day Fill Volume, Night Fill Volume, or Last Fill Volume is programmed too
high. This can cause you to be overfilled if the volume is not appropriate
for your body's size.
For Tidal therapies, Total UF volume is programmed too low. This can
cause a gradual buildup of UF volume during the therapy.
Last Manual Drain is programmed to No, or the UF Target for the Last
Manual Drain is programmed too low. This can cause an incomplete last
Drain.
Stop
and
Go
are pressed during Tidal dwells over multiple dwell cycles.
This can reduce the volumetric accuracy of the device over the course of
successive Tidal Dwell cycles.
After a power failure during Prime, the
Go
button is pressed to start
therapy without closing all clamps first. This can cause a free flow of fluid
from one bag to another and/or to the patient during the time when LOAD
THE SET is displayed.
The door is opened during an alarm or System Error without closing all
clamps first. This can cause a free flow of fluid from one bag to another
and/or to the patient.
The transfer set is connected to the patient line before CONNECT
YOURSELF appears on the display screen. This can cause air to be
delivered to your peritoneal cavity, which can cause IIPV if you had fluid
in your peritoneal cavity prior to the Initial Drain.
At the start of Fill 1, the patient line clamp is opened after a Check Patient
Line alarm or Check Your Position alarm appears on the display screen
without first initiating a manual drain. This can cause air to be delivered
to your peritoneal cavity, which can cause IIPV if you had fluid in your
peritoneal cavity prior to the Initial Drain.
Go
is pressed at the end of therapy before all clamps are closed when
CLOSE ALL CLAMPS appears on the display screen. This can cause a free
flow of fluid from one bag to another and/or to the patient.
Содержание HomeChoice
Страница 11: ...Table of Contents viii HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 12: ...Section 1 Glossary 1 Glossary...
Страница 27: ...Section 2 User Assistance Information 2 User Assistance Information...
Страница 30: ...Section 3 Warnings and Cautions 3 Warnings and Cautions...
Страница 47: ...Section 4 Indications for Use 4 Indications for Use...
Страница 49: ...4 Indications for Use 4 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 50: ...Section 5 Description 5 Description...
Страница 67: ...Section 6 Environmental Conditions 6 Environmental Conditions...
Страница 70: ...Section 7 Setup and Check out 7 Setup and Check out...
Страница 75: ...Section 8 Operating Instructions PRO Card and Modem 8 PRO Card and Modem...
Страница 94: ...Section 9 Operating Instructions Change Program 9 Change Program...
Страница 157: ...Section 10 Operating Instructions Make Adjustments 10 Make Adjustments...
Страница 174: ...Section 11 Operating Instructions Prepare for Therapy 11 Prepare for Therapy...
Страница 201: ...Section 12 Operating Instructions Perform Therapy 12 Perform Therapy...
Страница 223: ...12 Operating Instructions Perform Therapy 12 22 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 224: ...Section 13 Operating Instructions End Therapy 13 End Therapy...
Страница 237: ...Section 14 Operating Instructions Effluent Sampling 14 Effluent Sampling...
Страница 244: ...Section 15 Cleaning 15 Cleaning...
Страница 248: ...15 Cleaning 15 4 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 249: ...Section 16 Maintenance 16 Maintenance...
Страница 251: ...16 Maintenance 16 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 252: ...Section 17 Storage 17 Storage...
Страница 254: ...17 Storage 17 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 255: ...Section 18 Troubleshooting 18 Troubleshooting...
Страница 323: ...18 Troubleshooting 18 68 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 324: ...Section 19 Technical Data 19 Technical Data...
Страница 345: ...Section 20 Quick Reference 20 Quick Reference...
Страница 368: ...Index Index...