18. Troubleshooting
18-16
HomeChoice
APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
LOW DRAIN VOLUME
Display Message:
Cause:
A Slow Flow, No Flow, or air and fluid condition
occurred before the programmed Minimum Drain
Volume % (or Initial Drain Volume) completed.
A No Flow condition occurred before the Low Fill Mode
Minimum Drain Time completed.
This is an Auto Restart Alarm.
To Correct:
1. Change your position to try to drain more fluid.
2. Check for kinks in your patient line.
3. Correct the problem you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP
to mute the alarm.
2. Change your position or lower the cycler by six (6)
inches.
3. Press
GO
to return to your therapy.
If the alarm continues during your therapy, or if the alarm
occurs regularly over multiple therapies:
1. Follow the steps below to check your Drain Volume.
2. Contact your dialysis center to learn when it is safe to
LOW DRAIN VOLUME
Содержание HomeChoice
Страница 11: ...Table of Contents viii HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 12: ...Section 1 Glossary 1 Glossary...
Страница 27: ...Section 2 User Assistance Information 2 User Assistance Information...
Страница 30: ...Section 3 Warnings and Cautions 3 Warnings and Cautions...
Страница 47: ...Section 4 Indications for Use 4 Indications for Use...
Страница 49: ...4 Indications for Use 4 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 50: ...Section 5 Description 5 Description...
Страница 67: ...Section 6 Environmental Conditions 6 Environmental Conditions...
Страница 70: ...Section 7 Setup and Check out 7 Setup and Check out...
Страница 75: ...Section 8 Operating Instructions PRO Card and Modem 8 PRO Card and Modem...
Страница 94: ...Section 9 Operating Instructions Change Program 9 Change Program...
Страница 157: ...Section 10 Operating Instructions Make Adjustments 10 Make Adjustments...
Страница 174: ...Section 11 Operating Instructions Prepare for Therapy 11 Prepare for Therapy...
Страница 201: ...Section 12 Operating Instructions Perform Therapy 12 Perform Therapy...
Страница 223: ...12 Operating Instructions Perform Therapy 12 22 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 224: ...Section 13 Operating Instructions End Therapy 13 End Therapy...
Страница 237: ...Section 14 Operating Instructions Effluent Sampling 14 Effluent Sampling...
Страница 244: ...Section 15 Cleaning 15 Cleaning...
Страница 248: ...15 Cleaning 15 4 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 249: ...Section 16 Maintenance 16 Maintenance...
Страница 251: ...16 Maintenance 16 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 252: ...Section 17 Storage 17 Storage...
Страница 254: ...17 Storage 17 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 255: ...Section 18 Troubleshooting 18 Troubleshooting...
Страница 323: ...18 Troubleshooting 18 68 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Страница 324: ...Section 19 Technical Data 19 Technical Data...
Страница 345: ...Section 20 Quick Reference 20 Quick Reference...
Страница 368: ...Index Index...