b.
Press the key for the infusion pa-
rameter you wish to modify (
PRI VTBI
or
PRI RATE
).
c.
Enter the new value on the keypad.
d.
Press the
NEXT
key to enter the new
value.
e.
Modify additional PDP steps if de-
sired. After reviewing or changing
the last step, press
NEXT
again, until
the first step is displayed.
f.
To begin the PDP, press
PRI START
.
The pump begins delivering fluid.
6.
To program a new PDP:
a.
After completing steps 1 through 3
above, enter the desired values us-
ing one of the following methods:
•
Entering rate and VTBI—
Use
PRI RATE
,
PRI VTBI
and the nu-
meric keypad exactly as if pro-
gramming a primary infusion.
You may enter rate and volume
in either order.
•
Entering
VTBI
and
time—
First press
PRI VTBI
and
enter a value. Then press
TIME
and enter a time span for this
step (up to 99 hours and 99 min-
utes). Finally, press
PRI RATE
.
The device calculates and dis-
plays the resulting flow rate. If
the display reads
Hi
or
Lo
, the
calculated rate is outside the de-
vice's capabilities. Re-enter the
VTBI and the time and press
PRI
RATE
again.
•
Entering rate and time—
First
press
PRI RATE
and enter a value.
24
7-19-B1-564
Содержание Flo-Gard 6301