Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 70
53007410 Rev. C
CAUTION
A calibration bag must be used during all priming/verifying and Universal
Ingredient flushes.
2.
If a calibration bag is not already attached, attach it. Refer to Attaching the Calibration
3.
At the
Calibrating pump
message, tap
OK
.
Message
4.
If the
Bag currently on the load cell does not appear to be empty
message appears,
visually check the contents of the bag.
•
If the bag is empty, refer to The bag currently on the load cell does not appear to be
•
If the bag contains fluid, continue with the next step.
5.
Check whether the bag has space for an additional 200 mL of fluid.
•
If the bag does not have space:
a.
Tap
No
.
b.
At the
Operation Cancelled
message, tap
OK
.
c.
Remove the calibration bag. Refer to Removing the Calibration Bag on Page 73.
d.
Attach an empty calibration bag. Refer to Attaching the Calibration Bag on
Page 73.
e.
Return to Step 1.
•
If the bag has space, tap
Yes
.
IMPORTANT!
It is acceptable to calibrate the pump when the calibration bag
contains fluid. However, the calibration procedure is the only time you should tap
Yes
at this message. Never tap
Yes
when compounding a solution into a patient bag.
Message
Содержание ExactaMix 2400 Compounder
Страница 2: ......
Страница 170: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 162 53007410 Rev C Sample MixCheck Report...
Страница 177: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 169 Sample Formula Report...
Страница 180: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 172 53007410 Rev C Sample Log Report...
Страница 185: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 177 Sample Formulary Report...
Страница 199: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 191 Sample Bag Usage Report...