Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 69
If a cosigner was logged in, the software automatically logs out the cosigner and logs in
the original user.
If you started this procedure from the
Setup Wizard
screen, a check mark now appears
next to
Prime and Verify
at the
Setup Wizard
screen.
CALIBRATING THE COMPOUNDER
This procedure calibrates the compounder’s pump to ensure that it delivers the intended
volume of each ingredient.
1.
At the
Setup Wizard
screen, tap
Calibrate Compounder
.
Setup Wizard screen, calibrating the compounder
NOTE:
You can calibrate the compounder at any time by tapping
Compound > Calibrate
Pump
at the menu screen.