7.
While continuing to hold the inlet near the source container:
a.
Check that the number on the inlet label matches the port number.
b.
Check that the product attached to the inlet matches the information on the
screen.
c.
At the ingredient detail window, tap
Prime
.
d.
Check that a calibration bag is attached.
e.
At the
Priming pump
message, tap
OK
.
Message
f.
Watch and feel for fluid moving through the inlet that is in your hand.
When an ingredient is being primed, its button becomes yellow. The screen
displays an animation of the process.
PRIME AND VERIFY screen, priming in progress