25
Figure 4.
Gas-Rover with top of case and main circuit board removed.
CAUTION:
After replacing a methane or a CO sensor, perform the New Filament procedure
(described under Part 2, Setup and Calibration) and then completely calibrate
the instrument (both with 2.5% and 100% gas).