background image

BARTINGTON INSTRUMENTS

 

Page 17 of 19 OM2441/1

9. Care and Maintenance

The SCU1 requires no routine maintenance. The only user serviceable parts are the fuses in the 
IEC mains inlet socket.

9.1. Fuses

To change a blown fuse, ensure the mains supply is disconnected, then extract the fuse block 
from the IEC socket with a blunt tool (e.g. flat-bladed screwdriver). Extract and replace the blown 
fuse. Replace the fuse block.

WARNING:

 Fuses fitted must be of the type specified in the product brochure.

9.2. Calibration

Return the SCU1 to Bartington Instruments for calibration at the recommended intervals. Refer 
to the Calibration Certificate for further details.

9.3. Cleaning

WARNING:

 Disconnect the electrical supply before performing any cleaning operation.

Periodic cleaning is not normally required.

If the system becomes soiled and cleaning is necessary:

1.  Use a damp cloth to clean the outer surfaces.

2.  Use an air duster to blow debris from the connectors.

WARNING:

 Ensure water does not enter the system. The system must be completely dry 

before the electrical supply is reconnected.

Caution:

 Never use chemicals, such as solvents, when cleaning the SCU1.

Caution:

 Take particular care when cleaning around electrical connections. Bent or 

damaged pins may cause the magnetometer to malfunction.

Содержание SCU1

Страница 1: ...Operation Manual for SCU1 Signal Conditioning Unit...

Страница 2: ...1 Back Panel Connections and Controls 7 5 2 Front Panel Displays and Controls 8 6 Installing the SCU1 10 6 1 Location of the Equipment 10 6 1 1 Environmental Precautions 10 6 1 2 Mounting 10 6 1 3 Ori...

Страница 3: ...Controls 13 7 4 2 Gain Controls 13 7 4 3 Low Pass Filter LPF 14 7 4 4 High Pass Filter HPF 14 7 5 Using the SCU1 Outputs and Displays 15 7 5 1 Unconditioned and Conditioned Outputs 15 7 5 2 Display Pa...

Страница 4: ...aution is ignored Identifies items that must be disposed of safely to prevent unnecessary damage to the environment Note Provides useful supporting information on how to make better use of your purcha...

Страница 5: ...via the following outputs unconditioned analogue outputs signals direct from the magnetometer conditioned analogue signals the magnetometer signals modified by the SCU1 analogue conditioning digital d...

Страница 6: ...t 3 Gain Finally the DC adjusted output signal can be amplified by a variable gain stage to produce the conditioned output signal Offset and gain are controlled independently for the X Y and Z channel...

Страница 7: ...ter Output Type Selector Switch The SCU1 can be used with magnetometers that have unbalanced single ended outputs or balanced differential output signals This switch sets the SCU1 to operate in unbala...

Страница 8: ...HPF Control The SCU1 can apply high pass filtering to the magnetometer output signals This control sets the 3dB point frequency for the HPF 9 Magnetometer Supply Voltage Control This control sets the...

Страница 9: ...et voltage is subtracted Table 1 Offset switch functions 12 Offset Coarse Controls one for each channel 13 Offset Fine Controls one for each channel These rotary controls set the offset DC voltages ap...

Страница 10: ...ckets 6 1 3 Orientation The SCU1 can be orientated horizontally or vertically 6 1 4 Temperature To minimise temperature induced drift effects position the SCU1 in a constant ambient temperature out of...

Страница 11: ...ometer Supply Voltage Before switching the equipment ON set the appropriate voltage for your magnetometer using the front panel control Figure 2 item 9 For all Mag648 and Mag 03RC magnetometers use th...

Страница 12: ...ometer with a range of 60 T and an analogue output of 3V has a scale factor of 20 To convert voltage to field strength 1 Divide the output voltage by the gain control setting then 2 Multiply by the sc...

Страница 13: ...e adjust to set an approximate level and a fine adjust to achieve the exact offset level desired A polarity switch selects whether the offset is added to or subtracted from the magnetometer output or...

Страница 14: ...ing AC fields or very slow changes to a background DC field In these situations any DC signal present would normally be removed by using offset This allows the level of gain to be set as high as possi...

Страница 15: ...C content of the signals need to be measured e g with an oscilloscope meter or an A to D system or when a voltage output is required for connection for example to a feedback or control system Note The...

Страница 16: ...artington com or telephone the Bartington Instruments service team on 44 0 1993 706565 Fault Possible cause Solution Panel meters are off Power fault Check SCU1 is connected to a mains supply Check th...

Страница 17: ...truments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 9 3 Cleaning WARNING Disconnect the electrical supply before performing any cleaning oper...

Страница 18: ...NSTRUMENTS Page 18 of 19 OM2441 1 10 End of Life Disposal This product electrical and electronic equipment should not be placed in municipal waste Check local regulations for disposal of electronic pr...

Страница 19: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Отзывы: