6.2. Connecting the Equipment

Connect the equipment using the following sequence:

1.  Connect the magnetometer to the SCU1 magnetometer socket. Ensure the connector pins 

are correctly aligned with those in the socket. The locking ring should be hand-tightened 
only.

2.  If your system will use the SCU1 analogue outputs, connect the conditioned and/or 

unconditioned BNC outputs to your external equipment, as required.

3.  Ensure the IEC socket mains switch (Figure 1, item 5) is OFF (position 0). Connect your 

mains supply cable to the IEC socket. The unit will operate with a mains supply in the range 
specified in the product brochure. Range selection is automatic within the unit.

6.3. Initial Settings

6.3.1. Magnetometer Supply Voltage

Before switching the equipment ON, set the appropriate voltage for your magnetometer, using 
the front panel control (Figure 2, item 9).

For all Mag648 and Mag-03RC magnetometers, use the 12V setting.

For Mag-03MC, MS and IE magnetometers, the voltage setting should be used to compensate for 
any voltage loss in the magnetometer cable. Refer to individual sensor brochures to select the 
correct setting for your cable length (see the Mag-03 product brochure DS0013, available from 
the Bartington Instruments website, for recommended cable types).

6.3.2. Magnetometer Output Type Selection

Before switching on the equipment, set the magnetometer output type selector switch (Figure 
1, item 3) to the correct position for your magnetometer: see Table 2. See individual sensor data 
sheets for those not mentioned in this table.

Magnetometer

Output Type

Mag-03MS

Unbalanced

Mag-03MC

Unbalanced

Mag-03IE

Unbalanced

Mag-03RC

Balanced

Mag648

Balanced

Содержание SCU1

Страница 1: ...Operation Manual for SCU1 Signal Conditioning Unit...

Страница 2: ...1 Back Panel Connections and Controls 7 5 2 Front Panel Displays and Controls 8 6 Installing the SCU1 10 6 1 Location of the Equipment 10 6 1 1 Environmental Precautions 10 6 1 2 Mounting 10 6 1 3 Ori...

Страница 3: ...Controls 13 7 4 2 Gain Controls 13 7 4 3 Low Pass Filter LPF 14 7 4 4 High Pass Filter HPF 14 7 5 Using the SCU1 Outputs and Displays 15 7 5 1 Unconditioned and Conditioned Outputs 15 7 5 2 Display Pa...

Страница 4: ...aution is ignored Identifies items that must be disposed of safely to prevent unnecessary damage to the environment Note Provides useful supporting information on how to make better use of your purcha...

Страница 5: ...via the following outputs unconditioned analogue outputs signals direct from the magnetometer conditioned analogue signals the magnetometer signals modified by the SCU1 analogue conditioning digital d...

Страница 6: ...t 3 Gain Finally the DC adjusted output signal can be amplified by a variable gain stage to produce the conditioned output signal Offset and gain are controlled independently for the X Y and Z channel...

Страница 7: ...ter Output Type Selector Switch The SCU1 can be used with magnetometers that have unbalanced single ended outputs or balanced differential output signals This switch sets the SCU1 to operate in unbala...

Страница 8: ...HPF Control The SCU1 can apply high pass filtering to the magnetometer output signals This control sets the 3dB point frequency for the HPF 9 Magnetometer Supply Voltage Control This control sets the...

Страница 9: ...et voltage is subtracted Table 1 Offset switch functions 12 Offset Coarse Controls one for each channel 13 Offset Fine Controls one for each channel These rotary controls set the offset DC voltages ap...

Страница 10: ...ckets 6 1 3 Orientation The SCU1 can be orientated horizontally or vertically 6 1 4 Temperature To minimise temperature induced drift effects position the SCU1 in a constant ambient temperature out of...

Страница 11: ...ometer Supply Voltage Before switching the equipment ON set the appropriate voltage for your magnetometer using the front panel control Figure 2 item 9 For all Mag648 and Mag 03RC magnetometers use th...

Страница 12: ...ometer with a range of 60 T and an analogue output of 3V has a scale factor of 20 To convert voltage to field strength 1 Divide the output voltage by the gain control setting then 2 Multiply by the sc...

Страница 13: ...e adjust to set an approximate level and a fine adjust to achieve the exact offset level desired A polarity switch selects whether the offset is added to or subtracted from the magnetometer output or...

Страница 14: ...ing AC fields or very slow changes to a background DC field In these situations any DC signal present would normally be removed by using offset This allows the level of gain to be set as high as possi...

Страница 15: ...C content of the signals need to be measured e g with an oscilloscope meter or an A to D system or when a voltage output is required for connection for example to a feedback or control system Note The...

Страница 16: ...artington com or telephone the Bartington Instruments service team on 44 0 1993 706565 Fault Possible cause Solution Panel meters are off Power fault Check SCU1 is connected to a mains supply Check th...

Страница 17: ...truments for calibration at the recommended intervals Refer to the Calibration Certificate for further details 9 3 Cleaning WARNING Disconnect the electrical supply before performing any cleaning oper...

Страница 18: ...NSTRUMENTS Page 18 of 19 OM2441 1 10 End of Life Disposal This product electrical and electronic equipment should not be placed in municipal waste Check local regulations for disposal of electronic pr...

Страница 19: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Отзывы: