BARTINGTON INSTRUMENTS
Page 11 of 18 OM2443/4
Note:
If the battery voltage falls below 5V a warning buzzer will sound when the PSU1 is
switched on.
Note
: Battery capacity will decrease over time, depending on age and use.
Caution:
The operating temperature range is altered during battery charging.
6.3. Location of the Equipment
6.3.1.
Potentially Hazardous Locations
WARNING:
The charger supplied with this equipment is powered by mains electricity.
Do not use in wet or damp locations, where water may enter the unit and create a safety
hazard.
6.3.2.
Temperature
To minimise temperature induced drift effects, position the PSU1:
• in a constant ambient temperature
• out of direct sunlight.
6.3.3.
Proximity to Other Equipment
The PSU1 contains no high frequency electronics likely to cause emissions which could create
interference with other equipment. The unit is unlikely to be affected by interference from other
equipment in the normal operating environment.
Note
: The PSU1 is built with ferromagnetic materials and should therefore be kept at least
1 metre away from any magnetometers.
Caution:
The PSU1 should not be located in an environment with high radio frequency
noise as this would interfere with BNC connection to a DVM or Acquisition unit.