BARTINGTON INSTRUMENTS
Page 17 of 20 OM2448/4
Magmeter-2 stuck
in Balanced or
Unbalanced mode
Defective component
Contact Bartington service team regarding
repair
One or both LEDs do
not light correctly
Defective component
Contact Bartington service team regarding
repair
Backlight does not
illuminate the screen
Button not pressed and
held
Press and hold button to maintain backlight
Magmeter-2 defective
Contact Bartington service team regarding
repair
With a sensor
connected, all output
signals are faulty
Coupled incorrectly
Check position of AC/DC coupling switch
Balance mode set
incorrectly
Check position of balanced/unbalanced
mode switch
Defective sensor cable
Check sensor cable using relevant datasheet.
If found to be defective, contact Bartington
service team regarding repair
Sensor defective
Contact Bartington service team regarding
repair
Magmeter-2 defective
Contact Bartington service team regarding
repair
With a sensor
connected, one or
two output signals are
faulty
Defective sensor cable
Check sensor cable using relevant datasheet.
If found to be defective, contact Bartington
service team regarding repair
Sensor defective
Contact Bartington service team regarding
repair
Magmeter-2 defective
Contact Bartington service team regarding
repair
No output displayed
(when not mains
connected via the plug-
in charger adaptor)
Batteries flat
Recharge the batteries
Batteries expired
Replace batteries and charge
Batteries not fitted
Fit batteries and charge
None of the above
Contact Bartington service team regarding
repair