1.  Use a water-blocked cable.

2.  Reduce the risk of stress on the soldering by ensuring there is adequate slack in the wire 

between the cable and the wire terminations.

3.  Brace the slack with epoxy resin before moulding the cable to the connector.

4.  Fit additional protection in the form of a plastic hose, or sleeve, around the cable(s) at the 

emergence point.

Note:

 Cables are particularly prone to wear and damage at the point where they emerge at 

the surface of the ground, or sea.

4.3. Pre-Installation Tests

Prior to the installation of the system, the magnetometer, cable and power supply must be fully 
tested to ensure correct function as follows:

Caution: 

Take care to avoid bending, or otherwise damaging the contacts whilst conducting 

the tests.

1.  Test the cables for continuity (using an electrical continuity tester or ohmmeter).

a.  Test the cables end to end at the connectors to ensure the correct pins have been 

allocated to the conductors and there are no open, or high resistance circuits.

Note:

 Cable resistance will vary: refer to the product brochure for the expected values.

b.  Test the cables at the connectors to ensure there are no short circuits between the 

conductors.

2.  Check the power supply output voltage using a voltmeter. Refer to the product brochure 

for the required values.

Caution: 

Ensure the polarity is correct. Incorrect polarity is likely to irreparably damage 

the sensor.

Note:

 Bartington recommends the use of a current-limited power supply.

3.  Connect the magnetometer to the cable connector.

4.  Connect the power supply to the other cable connector.

5.  Switch on the power supply and wait until the magnetometer has stabilised. Refer to the 

product brochure for warm-up times.

6.  Confirm no magnetic objects are moving in the vicinity.

Содержание Mag648

Страница 1: ...Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Страница 2: ...e Magnetometer Environment Recommendations 5 4 2 Cable Recommendations 5 4 3 Pre Installation Tests 6 4 4 Mounting Recommendations 7 4 5 Post Installation Testing 7 5 Using the Mag648 and Mag649 8 5 1...

Страница 3: ...be disposed of safely to prevent unnecessary damage to the environment Note A note provides useful supporting information and sometimes suggests how to make better use of your purchase 2 Safe Use WARN...

Страница 4: ...ltages in response to the magnitude and direction of the orthogonal components of a magnetic field A right hand co ordinate system is adopted see figure 1 In this system the X Y and Z axis correspond...

Страница 5: ...d that such an evaluation be carried out using total field or resonance magnetometers 4 2 Cable Recommendations The standard magnetometer provides differential output lines for analogue signal transmi...

Страница 6: ...the cables for continuity using an electrical continuity tester or ohmmeter a Test the cables end to end at the connectors to ensure the correct pins have been allocated to the conductors and there ar...

Страница 7: ...to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Note Other versions of the Mag648 and Mag649 have plug in type...

Страница 8: ...Precautions Refer to the product brochure for maximum environmental electrical and mechanical ratings Caution Exceeding the maximum environmental ratings may cause irreparable damage to your sensor 6...

Страница 9: ...elpdesk for further advice If repair is not advised replace with a new unit For information about disposal of the damaged unit refer to End of Life Disposal 7 Care and Maintenance Unpackaged versions...

Страница 10: ...en to dispose of your magnetometer refer to the section on Troubleshooting 8 1 Waste Electrical and Electronic Equipment WEEE Regulations This product electrical and electronic equipment should not be...

Страница 11: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Отзывы: