HELMHOLTZ COIL INSTALLATION AND CALIBRATION
Z pK-pK input Scale = 1.0
X DUT gain increace factor = 1.2
Y DUT gain increace factor = 1.2
Z DUT gain increace factor = 1.2
Noise Input Gain = 1.3
DUT gain increace Label = 1.2
DUT PSU Scale Factor = 5.000000E-1
DUT Current Monitor Scale Factor = 1.0
DUT + Current Monitor Offset = 0E-6
DUT - Current Monitor Offset = 0E-6
DUT Voltage Monitor Scale Factor = 2.0
X Noise Offset = 1.0
Y Noise Offset = 1.0
Z Noise Offset = 1.0
(end)
X Coil Trim
Factor
Y Coil Trim
Factor
Z
Coil
Trim
Factor
Max coil field
HC1
5.3
5.3
5.3
5.000000E-4
HC2
7.3
6.3
7.3
2.000000E-4
HC9 or HC16
1.3
1.3
1.3
1.000000E-3
The final file is the sensor file *.tdf, which can be again created in notepad.
o
Mag03_100.tdf
(Start)
[Device]
Device Name = "Mag03-100"
(it is recommended to change the final number to match the sensor’s
full range)
Scale Factor = 10E-6 (may vary depending on sensor range
–
see table below)
Range = 100E-6 (may vary depending on sensor range
–
see table below)
Defaut Test Frequecy = 1
Default sample Rate = 10000
Default Number Of Samples = 50000
Device Supply Voltage = 12
Input Type = "Differential"
Default Input Filter = 10000
Test Results Path =
“/C/Bartington/TR/
Test results.trp"
[Alignment Error]