P-600 SIC Analyzer Manual
Chapter III: Installation and Startup
17
The P-600 Salt in Crude Analyzer requires an independent 110-120 or 230-240VAC (±10%), 50/60Hz
power supply. The system is jumper selectable 110-120/220-240VAC (±10%) 50/60 Hz, single phase, 5A.
AC power connections are made through the Customer Connections box attached to the upper left of the
top enclosure (see Figure 3-2).
CAUTION: It is the
installer’s responsibility to verify that the jumpers
on
HD2 (798112 Rev. B) in Customer Connection Box are configured
properly for the supply voltage (see Figure 3-6).
Figure 3-5: Customer Connections Enclosure
Figure 3-6: Voltage selection header (shown 220VAC, pin2-3)
CONTROL ROOM SIGNALS
The Analyzer incorporates analog, serial, and relay output capability. Modbus output may be provided as
a factory installed option.