11-07-2013
P-800LT Freeze Point Analyzer
6-4
Routine Maintenance & Service
T
ROUBLESHOOTING
Problem / Symptom
Cause
Corrective Action
Freeze point out of
range
Detected freeze point value is
outside the expected range of the
Analyzer
Faulty temperature sensor
Incorrect 4-20 mA signal
Allow unit to run additional cycles.
Process fluctuations or changeovers
can cause out of range alarms due to
different sample characteristics.
Check and replace as required.
Check 4-20 mA signal and range;
correct as required (see Section 4).
Freeze point not
detected
Time to detect value too short
Faulty cryocooler
Increase time to detect interval (see
Section 4).
Check and replace as required.
Erratic readings
Faulty temperature sensor
Faulty RTD PCB
Electronics failure (main PCB)
Check and replace as required.
Check and replace as required.
Check and replace as required.
Cell won’t cool
Low or no coolant supply to
cryocooler
Faulty cryocooler
Sample solenoid leaking or faulty
Check and correct as required.
Check and replace as required.
Check for proper flow during analysis;
replace as required.
Low or no sample flow
Damaged or plugged back
pressure regulator
Sample solenoid leaking or faulty
Faulty air-actuated valve
Blockage in cooling cell
Clean or replace as required.
Check and replace as required.
Check and replace as required.
Check and clear blockage as required.
4-20 mA failed or out of
range
Incorrect 4-20 mA values
Faulty 4-20 module
Faulty main PCB
Faulty line to control room
Correct as required (see Section 4).
Check and replace as required.
Check and replace as required.
Check line and repair as required.
Содержание ORB P-800LT
Страница 32: ...11 07 2013 P 800LT Freeze Point Analyzer 3 14 Installation and Startup THIS PAGE INTENTIONALLY BLANK...
Страница 58: ...11 07 2013 P 800LT Freeze Point Analyzer 6 6 Routine Maintenance Service CONTROL ENCLOSURE...
Страница 59: ...11 07 2013 P 800LT Freeze Point Analyzer 6 7 Routine Maintenance Service CONTROL ENCLOSURE DOOR...
Страница 60: ...11 07 2013 P 800LT Freeze Point Analyzer 6 8 Routine Maintenance Service THIS PAGE INTENTIONALLY BLANK...