27
Parameters” at the beginning of this section,
follow the procedures outlined in the proceed-
ing sections of this manual.
Creating a New Program or Editing an
Existing Program (4x16 Segment
Programmable Models Only)
The same steps are used when creating a
new program and editing an existing program
with the exception being that a new program
starts with all its segments set to End in the
tYPE parameter. Temporary changes can be
made to these parameters when the program
is in the hold state but permanent changes
must be made in the reset state. Follow the
steps below to create or edit a program.
1.
Press the PAGE button until you
reach the program list (ProG LiSt).
2.
Press the SCROLL button until dis-
play reads, “PrG.n.”
3.
Press the UP or DOWN button to
select a number for a new program or
to edit an existing program.
Hb: Holdback
Holdback consists of a value and a type. If
the measured value lags behind the setpoint
by an undesirable amount during a ramp or
dwell, the holdback feature can be used to
freeze the program at its current state (the
HOLD light will flash). The program will
resume when the error comes within the hold-
back value.
OFF:
holdback is disabled.
8 S
EGMENT
& 4
X
16 S
EGMENT
P
ROGRAMMABLE
M
ODELS W
/OTP