8
Unpackaging and Installation
1.
Remove the unit from its shipping container. Ensure
that the Ventgard
®
reservoir cap and power cord are
removed from the packaging materials before discard-
ing.
EASYpure High purity/Low TOC Cartridge
D50229
Pretreatment Cartridge DI Feed
D50230
Pretreatment Cartridge R/O - Distilled Feed
D50231
Ultrapure Mixed Bed Cartridge
D50233
2.
Place the EASYpure RF on a bench top convenient to
your work area, noting the Clearance Requirements.
3.
Using 1/2" I.D. tubing and tubing connectors (user
supplied), connect the overflow drain tubing (lower
right-hand corner of the rear of the unit) to an atmo-
spherically vented sink or floor drain. If an atmospheri-
cally vented sink or floor drain is not accessible, a
bucket or other container may be placed on the floor
beneath the EASYpure RF’s location and emptied
when full.
NOTE
Cartridges and the 0.2 micron final filter are not
provided with the EASYpure RF and must be
ordered separately.
WARNING
Do not place the EASYpure RF directly over
equipment that requires electrical service. Rou-
tine maintenance of this unit may involve water
spillage and subsequent electrical shock hazard
if improperly located.
NOTE
Overflow will occur only as a result of overfilling
by the user.
Unpackaging and Installation