115 VAC to 230 VAC
Parts Required
•
Two 8 Amp time delay fuses (Part No. 264306)
•
Suitable 230 VAC plug or plug adapter
1.
On the Power I/O PC Board: Remove jumpers
from J5 to J7 and J3 to J8. Install jumper from
J7 to J8. (Save other jumper for future reconver-
sion to 115 VAC.)
2.
Move white heater wire from J11 to J12. Move
black heater wire from J14 to J11. Leave other
black wire connected.
3.
On the Printer Power PC Board (MC10 with
Printer Only): Remove jumpers from JMP1 and
JMP2 positions and install jumper in JMP3 posi-
tion. (Save other jumper for future reconversion
to 115 VAC.)
4.
Install the 8 Amp fuses in fuseholders.
5.
Reinstall the left side panel, rear access panel,
and pressure relief valve deflection cover.
6.
Use either a plug adapter or remove the power
cord plug and install a suitable 230 VAC plug.
7.
Tag the sterilizer as configured for 230 VAC.