background image

Cimarec

Stirring Hot Plates

OPERATION MANUAL

AND PARTS LIST

Model #

Top Plate Size

Voltage

SP131014 4x4 

100

SP131015 4x4 

120

SP131010-33 4x4 

220-240

SP131324

7x7 100

SP131325 7x7

120

SP131320-33 7x7 

220-240

SP131634 10x10 

100

SP131635 10x10 

120

SP131630-33 10x10 

220-240

LT1313X1 • 10/31/06

Содержание Cimarec series

Страница 1: ...L AND PARTS LIST Model Top Plate Size Voltage SP131014 4x4 100 SP131015 4x4 120 SP131010 33 4x4 220 240 SP131324 7x7 100 SP131325 7x7 120 SP131320 33 7x7 220 240 SP131634 10x10 100 SP131635 10x10 120...

Страница 2: ...Installation 10 Unpacking 10 Installation 10 Operation 11 Setting the Stirring Speed 11 Setting the Temperature 12 To Achieve Fast Heat Up of Large Volumes 12 Heating Small Volumes 12 Heating Metal Ve...

Страница 3: ...materials Not rated for use in hazardous atmospheres 2 Use caution when heating volatile materials top surface and element can reach the Flash Point Temperature of many chemicals These stirring hot p...

Страница 4: ...modify grounded power plug Use only properly grounded outlets to avoid shock hazard 8 Use appropriate hand and eye protection when handling hazardous chemicals 9 Gross weight of items placed on top of...

Страница 5: ...Weight lbs kg 11 0 lbs 5 0 kg 11 0 lbs 5 0 kg 11 0 lbs 5 0 kg Top Plate Width 7 25 18 4 cm 7 25 18 4 cm 7 25 18 4 cm Height 1 0 2 5 cm 1 0 2 5 cm 1 0 2 5 cm Depth 7 25 18 4 cm 7 25 18 4 cm 7 25 18 4 c...

Страница 6: ...ared camera mounted approx 26 above the center of the top plate surface temperature accuracy at the edges of the top plate may vary from the stated accuracy Stirring Speed Specifications Speed Range 6...

Страница 7: ...equipment for measurement control and laboratory use Part I General Requirements Safety EN 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use Part I Gener...

Страница 8: ...00 C on 10 x 10 models The temperature is controlled at the plate surface by an internal sensor The stirrer will accurately maintain stirring speeds from 60 rpm up to 1200 rpm The top plate on the Cim...

Страница 9: ...rring hot plate If the temperature measurement of the ceramic top is made with measurement devices other than infrared the error of the measuring technique may be greater than the error of the unit Th...

Страница 10: ...ckaging The following items are included in the shipment Cimarec Stirring Hot Plate Cord Stir Bar Adapter 100V models only Knob Operator s Manual If any of these items are missing from the carton cont...

Страница 11: ...To set the speed turn the STIR knob to your desired setting Turn the knob clockwise to increase the speed or counter clockwise to decrease the speed To discon tinue stirring turn knob counter clockwis...

Страница 12: ...of Large Volumes If you are heating larger volumes faster heating can be achieved by turning the heat control knob to maxi mum temperature until the solution starts to heat and then turning the setpo...

Страница 13: ...ature metal vessels and sand baths may be heated safely without the danger of the ceramic top breaking Use the lowest temperature setting possible for applications to limit thermal stress to the ceram...

Страница 14: ...General Cleaning Instructions Wipe exterior surfaces with lightly dampened cloth con taining mild soap solution 14 Maintenance...

Страница 15: ...d to thermocouple Failure of element circuit Replace Element Replace Control Board E11 Alternating current Failure to detect alternating Replace Control Board Call Tech Service measurement failure cur...

Страница 16: ...Exploded Views 4x4 Exploded View 4 6 7 9 8 10 12 13 14 11 16 15 Optional Accessory Ref A 5 1 2 3 220 240V Models Only 16...

Страница 17: ...17 EXPLODED VIEWS 17 19 Note routing of leads and thermocouple 20 Number located this side 21 18 4x4 Exploded View Top Plate...

Страница 18: ...EXPLODED VIEWS 1 2 3 220 240V Models Only 4 5 6 8 7 Ref A 9 10 11 12 13 14 15 Optional Accessory 16 7x7 Exploded View 18...

Страница 19: ...19 EXPLODED VIEWS 17 19 Note routing of leads and thermocouple 20 Number located this side 21 18 7x7 Exploded View Top Plate...

Страница 20: ...20 EXPLODED VIEWS 10x10 Exploded View 3 5 6 7 1 2 220 240V Models Only 8 9 11 Optional Accessory Ref A 10 14 12 13 15 16 4...

Страница 21: ...21 EXPLODED VIEWS 10x10 Exploded View Top Plate 17 19 Note routing of leads and thermocouple 20 Number located this side 21...

Страница 22: ...22 Wiring Diagram...

Страница 23: ...3X1 AY1313X1 AY1313X1 Retaining Clip 100V 120V 16 AY1313X2 AY1313X2 AY1313X2 Retaining Clip 220 240V 17 JNX33 JNX35 Lower Insulation 18 JNX34 JNX36 JNX38 Upper Insulation 19 ELX41 ELX44 ELX47 Heating...

Страница 24: ...this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the deal er cannot handle your request then contact...

Страница 25: ...25...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...use inconsistent with written operating instructions for the product mishandling con tamination overheating modification or alteration of the product by any customer or third party or iii use of repla...

Отзывы: