38
Site~Rite* 6
Ultrasound System
Recommended Separations Distances between portable and mobile RF Communications
equipment and the Site~Rite* 6 Ultrasound System
The Site~Rite* 6 Ultrasound System is intended for use in the electromagnetic environment in which radi-
ated disturbances are controlled. The customer or user of the Site-Rite* 6 Ultrasound System can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the Site~Rite* 6 Ultrasound System as recommended below, according to
the maximum output power of the communications equipment.
Max Output Power
(Watts)
Separation (m)
150 kHz to 80 MHz
D = 1.2( √ P)
Separation (m)
80 kHz to 800 MHz
D = 1.2( √ P)
Separation (m)
800 MHz to 2.5 GHz
D = 2.3( √ P)
0.01
.1166
.1166
.2333
0.1
.3689
.3689
.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333