Display settings
8
NioWatch
to this calibration. If no former calibration was found, default
factory settings are selected.
When you restart NioWatch, the last selected display function
is automatically selected.
The following display functions are available:
Name
Description
DICOM
Select a DICOM display function for most
medical viewing applications.
The DICOM function results in more visible
grayscales in the images.
Dynamic Gamma 2.47
This gamma function best matches the
behavior of a grayscale CRT display.
The ‘dynamic’ gamma function is shifted
to take into account the non-zero lumi-
nance of an LCD panel when driven with a
“black” signal.
Dynamic Gamma 2.2
This gamma function best matches the
behavior of a color CRT display.
The ‘dynamic’ gamma function is shifted
to take into account the non-zero lumi-
nance of an LCD panel when driven with a
“black” signal.
Gamma 2.2
Select this display function in case the dis-
play is to replace a CRT display with a
gamma of 2.2
Gamma 1.8
Select this display function in case the dis-
play is to replace a CRT display with a
gamma of 1.8
Linear
If you select Linear, the native panel
behavior will not be corrected.