R5913350 /02
Nio Fusion 12MP
34
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends QAWeb as the solution of choice for calibration and QA. QAWeb brings many benefits
such as centralized asset management, the ability to schedule tasks, remote management, automated
reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA and AAPM
TG18. That's why QAWeb Agent acts as the master for all supported displays from the moment it is installed
and running. QAWeb Agent will take over from Embedded QA and overwrite any settings which were applied
by Embedded QA.
4.18.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status
(status since last compliance check)
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
4.18.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Compliance Check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
Содержание Nio Fusion12MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 12130 User guide Nio Fusion 12MP ...
Страница 10: ...R5913350 02 Nio Fusion 12MP 10 Welcome ...
Страница 20: ...R5913350 02 Nio Fusion 12MP 20 Installation and setup ...
Страница 41: ...41 R5913350 02 Nio Fusion 12MP Cleaning your display 5 ...
Страница 43: ...43 R5913350 02 Nio Fusion 12MP Repackaging instructions 6 ...
Страница 47: ...47 R5913350 02 Nio Fusion 12MP Image 6 7 Repackaging instructions ...
Страница 48: ...R5913350 02 Nio Fusion 12MP 48 Repackaging instructions ...
Страница 65: ......
Страница 66: ...www barco com R5913350 02 2021 01 26 Barco NV President Kennedypark 35 8500 Kortrijk Belgium ...