4. Advanced operation
4.12 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the DI-
COM display function is selected. Therefore, please refer to "Display functions", page
20 to correctly set the display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light cor-
rection value into account. This value is determined by the selected reading room. Therefore, it is also
important to select a realistic reading room when enabling ALC. This can be done by following the instruc-
tions in "Reading rooms", page 22.
To enable/disable ALC:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Ambient Light
menu.
3. Enter the
Ambient Light Compensation
submenu.
4. Select
Enabled/Disabled
as desired and con
fi
rm.
4.13 Reading rooms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 20 to correctly set the display
function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-de
fi
ned reading rooms.
Each of these reading rooms are de
fi
ned by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to
take into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to "Ambient Light Compensation (ALC)", page 22 to enable ALC.
The available reading rooms for your Nio Color 5MP are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Of
fi
ce:
Corresponds to light conditions in of
fi
ce rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical view-
ing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1. Bring up the OSD main menu.
22
K5903146 NIO COLOR 5MP 24/05/2017
Содержание Nio Color 5MP MDNC-6121
Страница 1: ...Nio Color 5MP User Guide 5 8MP high bright color display MDNC 6121 K5903146 01 24 05 2017 ...
Страница 8: ...1 Welcome 6 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 18: ...3 Daily operation Cancel Standby IEC 60417 5009 16 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 32: ...4 Advanced operation 30 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 34: ...5 Cleaning your display 32 K5903146 NIO COLOR 5MP 24 05 2017 ...