4. Advanced operation
To set the luminance target:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration
menu.
3. Enter the
Luminance Target
submenu.
4. Set a luminance target value as desired and con
fi
rm.
The default, factory DICOM calibrated luminance value is available in the technical spec-
i
fi
cations table. The guaranteed backlight lifetime is valid for this setting.
4.9
Viewing modes
About viewing modes
The Nio Color 5MP can be used in two viewing modes:
•
Diagnostic:
This mode provides the full calibrated luminance and is intended for using the display for
diagnostic purposes.
•
Text:
In this mode, the luminance is reduced to approximately half of the luminance. This is intended
for using the display with of
fi
ce applications such as word processing.
Please note that text mode is not persistent, once powered off, the unit will restart in diagnostic mode.
To quickly switch the viewing mode without having to enter the OSD menu, touch the
left and right key at the same time during normal operation.
The diagnostic mode should always be selected when the Nio Color 5MP is intended to
be used in a diagnostic environment.
To select a viewing mode:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration
menu.
3. Enter the
Viewing Mode
submenu.
4. Select
Diagnostic/Text
as desired and con
fi
rm.
4.10 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been de
fi
ned. These functions emphasize on these parts containing crucial diag-
nostic information by correcting the native panel behavior.
20
K5903146 NIO COLOR 5MP 24/05/2017
Содержание Nio Color 5MP MDNC-6121
Страница 1: ...Nio Color 5MP User Guide 5 8MP high bright color display MDNC 6121 K5903146 01 24 05 2017 ...
Страница 8: ...1 Welcome 6 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 18: ...3 Daily operation Cancel Standby IEC 60417 5009 16 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 32: ...4 Advanced operation 30 K5903146 NIO COLOR 5MP 24 05 2017 ...
Страница 34: ...5 Cleaning your display 32 K5903146 NIO COLOR 5MP 24 05 2017 ...