![Barco Nio 5MP LED Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/barco/nio-5mp-led/nio-5mp-led_user-manual_4112725025.webp)
4. Advanced operation
4.13.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure
the necessary luminance levels for either a calibration or compliance test. Various settings for both actions
can be selected from the display’s OSD menu. The last results of both actions can be consulted from the
OSD.
Embedded QA or MediCal QAWeb?
Embedded QA is not a replacement for the Barco MediCal QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb
brings many bene
fi
ts such as centralized asset management, the ability to schedule tasks, remote man-
agement, automated reporting, alerting and speci
fi
c support of regional QA standards such as DIN 6868-
57, JESRA and AAPM TG18. That’s why MediCal QAWeb Agent acts as the master for all supported dis-
plays from the moment it is installed and running. MediCal QAWeb Agent will take over from Embedded
QA and overwrite any settings which were applied by Embedded QA.
4.13.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status (status since last compliance check):
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation com-
pared to a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance
check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status:
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown:
Backlight runtime elapsed since latest calibration, display function, ambient light compensation, read-
ing room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration > Embedded QA
menu.
3. Select
DICOM status report
to make the information visible on the screen.
K5902121 NIO 5MP LED DISPLAY 18/01/2018
23
Содержание Nio 5MP LED
Страница 1: ...Nio 5MP LED Display User Guide MDNG 5221 K5902121 01 18 01 2018...
Страница 8: ...1 Welcome 6 K5902121 NIO 5MP LED DISPLAY 18 01 2018...
Страница 18: ...3 Daily operation 16 K5902121 NIO 5MP LED DISPLAY 18 01 2018...
Страница 34: ...5 Repacking instructions 32 K5902121 NIO 5MP LED DISPLAY 18 01 2018...
Страница 36: ...6 Cleaning your display 34 K5902121 NIO 5MP LED DISPLAY 18 01 2018...