K5902114 /03
Nio 3MP LED Display
27
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
Ambient Light Compensation (ALC)
, page 26 to enable ALC.
The available reading rooms for your Nio 3MP LED Display are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Reading Room
submenu.
4.
Select one of the available reading rooms and confirm.
4.14 Embedded QA
Overview
•
•
•
•
•
•
4.14.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display
’
s OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or MediCal QAWeb?
Embedded QA is not a replacement for the Barco MediCal QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb brings
many benefits such as centralized asset management, the ability to schedule tasks, remote management,
automated reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA and
AAPM TG18. That's why MediCal QAWeb Agent acts as the master for all supported displays from the
moment it is installed and running. MediCal QAWeb Agent will take over from Embedded QA and overwrite
any settings which were applied by Embedded QA.
Содержание Nio 3MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 3421 User Guide Nio 3MP LED Display...
Страница 5: ...5 K5902114 03 Nio 3MP LED Display Welcome 1...
Страница 16: ...K5902114 03 Nio 3MP LED Display 16 Installation...
Страница 17: ...17 K5902114 03 Nio 3MP LED Display Daily operation 3...
Страница 21: ...21 K5902114 03 Nio 3MP LED Display Advanced operation 4...
Страница 32: ...K5902114 03 Nio 3MP LED Display 32 Advanced operation...
Страница 33: ...33 K5902114 03 Nio 3MP LED Display Repacking your display 5...
Страница 36: ...K5902114 03 Nio 3MP LED Display 36 Repacking your display...
Страница 37: ...37 K5902114 03 Nio 3MP LED Display Cleaning your display 6...
Страница 39: ...39 K5902114 03 Nio 3MP LED Display Important information 7...
Страница 53: ...K5902114 03 Nio 3MP LED Display 53 Important information...