K5902114 /03
Nio 3MP LED Display
28
4.14.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status (status since last compliance check):
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status:
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown:
Backlight runtime elapsed since latest calibration, display function, ambient light compensation, reading
room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM status report
to make the information visible on the screen.
4.14.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM compliance check
to start the compliance check.
Warning:
Pressing a key during the compliance check will abort the check.
4.14.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
Содержание Nio 3MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 3421 User Guide Nio 3MP LED Display...
Страница 5: ...5 K5902114 03 Nio 3MP LED Display Welcome 1...
Страница 16: ...K5902114 03 Nio 3MP LED Display 16 Installation...
Страница 17: ...17 K5902114 03 Nio 3MP LED Display Daily operation 3...
Страница 21: ...21 K5902114 03 Nio 3MP LED Display Advanced operation 4...
Страница 32: ...K5902114 03 Nio 3MP LED Display 32 Advanced operation...
Страница 33: ...33 K5902114 03 Nio 3MP LED Display Repacking your display 5...
Страница 36: ...K5902114 03 Nio 3MP LED Display 36 Repacking your display...
Страница 37: ...37 K5902114 03 Nio 3MP LED Display Cleaning your display 6...
Страница 39: ...39 K5902114 03 Nio 3MP LED Display Important information 7...
Страница 53: ...K5902114 03 Nio 3MP LED Display 53 Important information...