K5902114 /03
Nio 3MP LED Display
45
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Nio 3MP LED Display
requires continued
operation during power
mains interruptions, it is
recommended that the Nio
3MP LED Display be
powered from an
uninterruptible power
supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
6
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385
380
–
390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430
–
470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704
–
787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2
0.3
9
745
6:
Nio 3MP LED Display doesn
’
t contain susceptible components to magnetic fields
Содержание Nio 3MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 3421 User Guide Nio 3MP LED Display...
Страница 5: ...5 K5902114 03 Nio 3MP LED Display Welcome 1...
Страница 16: ...K5902114 03 Nio 3MP LED Display 16 Installation...
Страница 17: ...17 K5902114 03 Nio 3MP LED Display Daily operation 3...
Страница 21: ...21 K5902114 03 Nio 3MP LED Display Advanced operation 4...
Страница 32: ...K5902114 03 Nio 3MP LED Display 32 Advanced operation...
Страница 33: ...33 K5902114 03 Nio 3MP LED Display Repacking your display 5...
Страница 36: ...K5902114 03 Nio 3MP LED Display 36 Repacking your display...
Страница 37: ...37 K5902114 03 Nio 3MP LED Display Cleaning your display 6...
Страница 39: ...39 K5902114 03 Nio 3MP LED Display Important information 7...
Страница 53: ...K5902114 03 Nio 3MP LED Display 53 Important information...