K5902157-01
Nio 3MP LED Display
26
•
Bluebase:
Simulation of the bluebase film color temperature.
•
User:
When selecting the User color temperature setting, you will be able to manually define the X and Y
coordinates or the display color temperature in separate submenus.
•
Native White:
The native, unmodified color temperature of the LCD panel.
To select a color preset:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Color Settings
menu.
3.
Enter the
Color Presets
submenu.
4.
Select one of the available Color Presets and confirm.
4.12 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the DICOM display
function is selected. Therefore, please refer to
“
”
, page 25 to correctly set the
display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
, page 26.
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.13 Reading rooms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected. Therefore,
please refer to
“
”
, page 25 to correctly set the display function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
Ambient Light Compensation (ALC)
, page 26 to enable ALC.
The available reading rooms for your Nio 3MP LED Display are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
Содержание Nio 3MP LED
Страница 5: ...5 K5902157 01 Nio 3MP LED Display Welcome 1...
Страница 8: ...K5902157 01 Nio 3MP LED Display 8 Welcome...
Страница 16: ...K5902157 01 Nio 3MP LED Display 16 Installation...
Страница 17: ...17 K5902157 01 Nio 3MP LED Display Daily operation 3...
Страница 21: ...21 K5902157 01 Nio 3MP LED Display Advanced operation 4...
Страница 32: ...K5902157 01 Nio 3MP LED Display 32 Advanced operation...
Страница 33: ...33 K5902157 01 Nio 3MP LED Display Repacking your display 5...
Страница 36: ...K5902157 01 Nio 3MP LED Display 36 Repacking your display...
Страница 37: ...37 K5902157 01 Nio 3MP LED Display Cleaning your display 6...
Страница 39: ...39 K5902157 01 Nio 3MP LED Display Important information 7...
Страница 51: ...K5902157 01 Nio 3MP LED Display 51 Important information...