28
4.
Select one of the available reading rooms and confirm.
4.16 Embedded QA
4.16.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display’s OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or QAWeb?
Embedded QA is not a replacement for the Barco QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends QAWeb as the solution of choice for calibration and QA. QAWeb brings many benefits
such as centralized asset management, the ability to schedule tasks, remote management, automated
reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA and AAPM
TG18. That's why QAWeb Agent acts as the master for all supported displays from the moment it is installed
and running. QAWeb Agent will take over from Embedded QA and overwrite any settings which were applied
by Embedded QA.
4.16.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status
(status since last compliance check)
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
•
Maximum error:
Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
•
Error threshold:
Shows the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
•
Time elapsed since latest compliance check:
Shows the backlight runtime since last compliance check.
•
Display Function:
Shows the current display function.
•
Ambient light compensation:
Shows the ambient light compensation status.
•
Reading Room:
Shows the selected reading room.
•
Luminance:
Shows the measured luminance.
•
Black luminance:
Shows the measured black luminance.
DICOM Calibration Status
•
No calibration executed yet:
No other information is visible
•
Calibration executed:
When the calibration is executed, the following extra information is shown: Time
elapsed since latest calibration, Display Function, Ambient Light Compensation and Reading Room.
Current DICOM Settings
•
Display Function:
Shows the current display function.
•
Ambient Light Compensation:
Shows the ambient light compensation status.
•
Reading room:
Shows the selected reading room.
To retrieve the DICOM status report:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Embedded QA
menu.
3.
Select
DICOM Status Report
to make the information visible on the screen.
Содержание MDNG-6221
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNG 6221 option NC EM User Guide 5 8MP high bright display Nio 5MP...
Страница 5: ...5 K5902142 04 Nio 5MP Welcome 1...
Страница 8: ...K5902142 04 Nio 5MP 8 Welcome...
Страница 9: ...9 K5902142 04 Nio 5MP Installation 2...
Страница 16: ...K5902142 04 Nio 5MP 16 Installation...
Страница 17: ...17 K5902142 04 Nio 5MP Daily operation 3...
Страница 34: ...K5902142 04 Nio 5MP 34 Advanced operation...
Страница 35: ...35 K5902142 04 Nio 5MP Cleaning your display 5...
Страница 37: ...37 K5902142 04 Nio 5MP Important information 6...
Страница 52: ...K5902142 04 Nio 5MP 52 Important information...
Страница 53: ......
Страница 54: ...www barco com K5902142 04 2022 01 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...