![Barco MDNG-6221 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/barco/mdng-6221/mdng-6221_user-manual_4112307026.webp)
K5902142 /04
Nio 5MP
26
3.
Enter the
Viewing Mode
submenu.
4.
Select
Diagnostic/Text
as desired and confirm.
To quickly switch the viewing mode without having to enter the OSD menu, touch the left and right
key at the same time during normal operation.
4.12 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been defined. These functions emphasize on these parts containing crucial diagnostic
information by correcting the native panel behavior.
The available display functions for your Nio 5MP are:
•
Native:
If you select Native, the native panel behavior will not be corrected.
•
Dynamic Gamma 1.8
or
2.2
: These are gamma functions that are shifted to take into account the non-zero
luminance of an LCD panel when driven with a “black” signal. They are especially useful in CT applications
to improve the perception of low Hounsfield values.
•
DICOM:
DICOM (Digital Imaging and Communications in Medicine) is an international standard that was
developed to improve the quality and communication of digital images in radiology. In short, the DICOM
display function results in more visible grayscales in the images. Barco recommends selecting the DICOM
display function for most medical viewing applications.
•
User:
This display function will be automatically selected when display functions are defined by QAWeb.
•
Gamma 1.8
or
2.2
: Select one of these display functions in case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
The settings of the display must be adapted to suit the requirements of the visualization software. In
case of doubt, please contact the vendor of the visualization software.
To select a display function:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Enter the
Display Function
submenu.
4.
Select one of the available display functions and confirm.
4.13 Color presets
About color presets
The available color preset settings for your display are:
•
Clearbase:
Simulation of the clearbase film color temperature.
•
Bluebase:
Simulation of the bluebase film color temperature.
•
User:
When selecting the User color temperature setting, you will be able to manually define the X and Y
coordinates or the display color temperature in separate submenus.
•
Native White:
The native, unmodified color temperature of the LCD panel.
To select a color preset:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Color Settings
menu.
Содержание MDNG-6221
Страница 1: ...ENABLING BRIGHT OUTCOMES MDNG 6221 option NC EM User Guide 5 8MP high bright display Nio 5MP...
Страница 5: ...5 K5902142 04 Nio 5MP Welcome 1...
Страница 8: ...K5902142 04 Nio 5MP 8 Welcome...
Страница 9: ...9 K5902142 04 Nio 5MP Installation 2...
Страница 16: ...K5902142 04 Nio 5MP 16 Installation...
Страница 17: ...17 K5902142 04 Nio 5MP Daily operation 3...
Страница 34: ...K5902142 04 Nio 5MP 34 Advanced operation...
Страница 35: ...35 K5902142 04 Nio 5MP Cleaning your display 5...
Страница 37: ...37 K5902142 04 Nio 5MP Important information 6...
Страница 52: ...K5902142 04 Nio 5MP 52 Important information...
Страница 53: ......
Страница 54: ...www barco com K5902142 04 2022 01 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...