7. Important information
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Power cords:
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
•
Do not overload wall outlets and extension cords as this may result in
fi
re or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to
be walked upon or pinched by items placed upon or against them, paying particular attention to cords
at plugs and receptacles.
•
The power supply cord should be replaced by the designated operator only at all time.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle
marked “Hospital Only” or “Hospital Grade” (located on product or power supply cord)
External equipment
External equipment intended for connection to signal input/output or other connectors, shall comply with
relevant UL/ EN/ IEC standard (e.g. UL/EN/IEC 60950 for IT equipment and ANSI/AAMI ES/EN 60601-1
/ IEC 60601 series for medical electrical equipment). In addition, all such combinations -systems- shall
comply with the standard IEC 60601-1, safety requirements for medical electrical systems. Equipment
not complying with ANSI/AAMI ES/EN / IEC 60601-1 shall be kept outside the patient environment, as
de
fi
ned in the standard.
Equipment not complying with IEC 60601 must be kept outside the patient environment, as de
fi
ned in the
standard as at least 1.5 meters from the patient or the patient support.
Any person who connects external equipment to signal input, signal output, or other connectors has formed
a system and is therefore responsible for the system to comply with the requirements of IEC 60601-1. If
in doubt, speak with a quali
fi
ed technician.
In locations where 240 V outlets are used, connect this display only on a center-tapped, 240 V, single-
phase supply.
Water and moisture
Never expose the display to rain or moisture.
Never use the display near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry
tub or in a wet basement.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a
cupboard or another closed location, heed the necessary space between the set and the sides of the
cupboard.
Installation
Place the device on a
fl
at, solid and stable surface that can support the weight of at least 3 devices. If you
use an unstable cart or stand, the device may fall, causing serious injury to a child or adult, and serious
damage to the device.
This apparatus conforms to:
CE (MDD 93/42/EEC class II b product), CE - 2014/30/EU, 2014/108/EC (directive 2004/108/EC
is repealed with effect from 20 April 2016), IEC 60601-1:2005+A1:2012, ANSI/AAMI ES60601-1
(2005/(R)2012 +A1:2012+C1:2009/(R)2012 + A2:2010/(R)2012) ,CAN/CSA-C22.2 No.
60601-1:14,
DEMKO - EN 60601-1:2006/A1:2013, GB9254-2008,GB4943.1-2011, GB17625.1-2012, KC, VCCI, FCC
class B, ICES-001 Level B, FDA 510 (k), RoHS.
32
K5903503 NIO 3MP LED DISPLAY 15/10/2014
Содержание MDNG-3220
Страница 1: ...NIO 3MP LED Display User Guide MDNG 3220 K5903503 01 15 10 2014...
Страница 4: ...Table of contents 2 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 10: ...2 Parts controls and connectors 8 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 18: ...3 Display installation 16 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 22: ...4 Daily operation 20 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 30: ...5 Advanced operation 28 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 32: ...6 Maintenance 30 K5903503 NIO 3MP LED DISPLAY 15 10 2014...
Страница 46: ...7 Important information 44 K5903503 NIO 3MP LED DISPLAY 15 10 2014...