R5913897 /00
MDMX-MNA
28
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MDMX-MNA is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDMX-MNA should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the MDMX-MNA requires
continued operation
during power mains
interruptions, it is
recommended that the
MDMX-MNA be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
1
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
1.
MDMX-MNA doesn
’
t contain susceptible components to magnetic fields
Содержание MDMX-22400
Страница 4: ...R5913897 00 MDMX MNA 4 ...
Страница 5: ...5 R5913897 00 MDMX MNA Welcome 1 ...
Страница 7: ...7 R5913897 00 MDMX MNA Image 1 1 MDMX 22400 Option GNTB Welcome ...
Страница 10: ...R5913897 00 MDMX MNA 10 Welcome ...
Страница 11: ...11 R5913897 00 MDMX MNA Parts controls and connectors 2 ...
Страница 14: ...R5913897 00 MDMX MNA 14 Image 2 4 Parts controls and connectors ...
Страница 15: ...15 R5913897 00 MDMX MNA Daily operation 3 ...
Страница 19: ...19 R5913897 00 MDMX MNA Cleaning and disinfection 4 ...
Страница 21: ...21 R5913897 00 MDMX MNA Important information 5 ...
Страница 38: ...R5913897 00 MDMX MNA 38 Important information ...
Страница 39: ......