5. Advanced operation
As the Coronis 5MP is intended to be used in a diagnostic environment, the diagnostic
mode should always be selected.
To select a viewing mode:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration
menu.
3. Enter the
Viewing Mode
submenu.
4. Select
Diagnostic/Text
as desired and con
fi
rm.
5.14 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been de
fi
ned. These functions emphasize on these parts containing crucial diag-
nostic information by correcting the native panel behavior.
Native
If you select Native, the native panel behavior will
not be corrected.
Dynamic Gamma 1.8
Dynamic Gamma 2.2
These are gamma functions that are shifted to take
into account the non-zero luminance of an LCD
panel when driven with a “black” signal. They are
especially useful in CT applications to improve the
perception of low Houns
fi
eld values.
DICOM
DICOM (Digital Imaging and Communications
in Medicine) is an international standard that
was developed to improve the quality and
communication of digital images in radiology.
In short, the DICOM display function results in
more visible grayscales in the images. Barco
recommends selecting the DICOM display function
for most medical viewing applications.
User
This display function will be automatically selected
when display functions are de
fi
ned by MediCal
QAWeb.
Gamma 1.8
Gamma 2.2
Select one of these display functions in case the
display is to replace a CRT display with a gamma
of 1.8 or 2.2 respectively.
To select a display function:
1. Bring up the OSD main menu.
2. Navigate to the
Con
fi
guration > Calibration
menu.
3. Enter the
Display Function
submenu.
4. Select one of the available display functions and con
fi
rm.
26
B4100592 CORONIS 5MP 05/08/2014
Содержание MDCG-5121
Страница 1: ...Coronis 5MP User Guide MDCG 5121 B4100592 03 05 08 2014...
Страница 4: ...1 Welcome 2 B4100592 CORONIS 5MP 05 08 2014...
Страница 8: ...2 Parts controls and connectors 6 B4100592 CORONIS 5MP 05 08 2014...
Страница 22: ...4 Daily operation 20 B4100592 CORONIS 5MP 05 08 2014...
Страница 34: ...5 Advanced operation 32 B4100592 CORONIS 5MP 05 08 2014...
Страница 36: ...6 Cleaning your display 34 B4100592 CORONIS 5MP 05 08 2014...