3. Display installation
-
1
: SDI
-
2
: DVI
-
3
: CVBS
-
4
: S-Video
1
2
3
4
Image 3-2
3.2
Power supply connection
To connect the power supply
1. Connect the supplied external DC power supply unit to the +24 VDC power input of your E192HSA
display.
2. Plug the other end of the power supply into a
grounded
power outlet by means of the proper power
cord delivered in the packaging.
Warning:
To avoid risk of electric shock, the external DC power supply must be connected to a mains
with protective earth. The ground connection on the display’s DC power input connector has
no protective earth function. The E192HSA display protective earth connection is provided
via a dedicate pin (see next steps).
Image 3-3
Protective earth
Earth the E192HSA by connecting the protective earth pin to a grounded outlet by means of an AWG18
wire (maximum admitted cable length according to national regulation requirements) and screw with lock
washer M4x12 max.
Image 3-4
12
K5902113
(
451920611383
)
E192HSA
27/11/2017