Dispensing Calibration
The parameters can not be changed under dispensing volume calibration
operation. The defaulted parameters are the parameters set under
Dispensing Setting.
Pump stops under dispense mode, press and turn
to highlight
calibration line. Press
to enter next interface. Turn
to select
Dispensing. Press
to enter next interface.
Press
, the LCD displays running time and actual volume.
Running time counts down.
Turn
to input actual volume. Press
for confirmation.
Knob
Knob
Knob
Knob
Run Key
Knob
Knob
If the actual volume is known in advance, enter the calibration interface
and input the actual volume directly.
Dispensing mode operation procedure
Exter nal Control
4
2 0 m A
.
1 0 V
~
~
~
~
0 .
5
5
V
1 0
1 . 0
1 0 k H z
External
Control Mode
(DB15 Plug)
Analog Control
Communication
Control
R 4 8 5
S
1.Enable the external control function.
2.These five Standard Analog Control Module need to be order separately.
Exter nal control setting
Set the pump to enable the external control or not.
On: Enable external control
Off: Disable external control
Set the external control
Press and turn
to highlight the external control line. Press
to
enter next interface, turn
to select On or Off, press
for
confirmation. Press
not to store the selection and return to
previous menu.
Knob
Knob
Knob
Knob
Exit Key
Analog Signal Control
Set “External Control” in “on” state. External control module is shown
as above.
Five kinds of standard external control module need to be ordered
separately according to special requirements.
9
10
Micro Gear Pump Drive Operating Manual
Micro Gear Pump Drive Operating Manual