1 . 4 .
Flow
:
7 . 7 0 5 m L / m i n
T i m e:1 3 . 2 3 "
Note:
Change the tubing if can't get the desired filling time.
1 .
1 . 1 .
2 0 : 4 7 : 4 9
I D 3 . 1
I N T
↑
1.800mL/min
1 . 4 .
:
Flow
7 . 7 0 5 m L / m i n
T i m e:1 3 . 2 3 "
1 .
2 0 : 4 7 : 4 9
I D 3 . 1
I N T
1.800mL/min
2 .
1 .
2 0 : 4 7 : 4 9
I D 3 . 1
I N T
1.800mL/min
2 .
2 . 1 .
S e t T e s t T i m e :
2 7 "
P r e s s
t o n e x t
2 . 1 .
2 . 1 .
C a l V o l:
3 0 m L
P r e s s
t o b e g i n
↑
Prompt:
In “Disp.” state, the testing time of flow rate calibration can be set the
time of disp. Time (please see the “Flow” setting of “Filling setting”) for
more accurate filling volume.
Enter “System Setting” interface.
There are 7 submenus in “System Setting” menu. They are “Tubing
ID”, “Remote Control”, “Pump I.D.”, “Footswitch”, “Back Suction”,
“Clock Setting” and “Pump Head”.
7
8
Running Interface
Running Interface
Running Interface
BT300-1F OPERATING MANUAL
BT300-1F OPERATING MANUAL
Fill Setting
Fill Setting
Fill Setting
Calibration
Calibration
System Setting
Turn the
Knob
Turn the
Knob
Turn the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Press the
Knob
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
Please make sure the pump
is in “Flow” or “Disp.” state.
See page 13 “Control Mode”
for setting method.
D i s p V o l
1 . 6 m L
a u s e T i m e
2 "
C o p y N o .
4
:
P
:
:
Flow: Adjust the dispensing and filling efficiency. Its range is
determined by the pump head and tubing.
In “Fill Setting” interface, press the
to enter the
next interface. Turn the
to highlight the “Flow”.
Press the
to enter “Flow” interface. Turn the
to adjust the flow rate and dispensing time.
Press the
to confirm, save and quit. Press the
to cancel the setting and return to the previous
interface.
: Filling time for one time.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Adjusting Knob
Exit Key
Time
Calibration Function
Calibration is needed in below situation:
1. Use the pump the first time; 2. Change the pump head; 3.
Change the tubing; 4. Setting the new flow rate; 5. Operating
condition changes.
Enter calibration function setting interface:
In “Running Interface”, press the
to enter “Fill
S e t t i n g ”
i n t e r f a c e ,
t u r n
t h e
t o
e n t e r
“Calibration” interface.
Adjusting Knob
A d j u s t i n g
K n o b
Flow Rate Calibration
In “Calibration” interface, press the
to enter “Test
Time” setting interface. Turn the
to set the test
time. Its range is 1s 999s. Press the
to enter the
next interface. Press the
to begin the testing.
Turn the
to input the actual volume after testing.
Press the
to confirm, save and quit. Press the
to cancel the calibration and return to the “Calibration”
interface.
Adjusting Knob
Adjusting Knob
Adjusting Knob
Start/Stop Key
Adjusting Knob
Adjusting Knob
Exit Key
A c t V o l:
3 1 m L
D i s p T i m e: 2 7 "
P r e s s
t o b e g i n
Press
Turn the Knob to adjust
flow rate and disp. time
Turn knob to
input actual the volume
Turn knob to set
test time