LEAD - 2 OPERATING MANUAL

LEAD - 2 OPERATING MANUAL

Optional Accessory

Communication Function

Maintenance

Warranty

Footswitch

Connect to the external control interface. Control the start and
stop of the pump.

Note:

Footswitch is optional accessories. It can only control the start
and stop of the pump. The speed and direction of the pump are
controlled by the knob and switch on the operation penal.

Lead-2 can connect to the control computer (computer, PLC, SCM)
through RS485 serial communication interface.

Note:

1.The pump I.D. of the pump which connects to the control

computer through RS485 serial communication interface must
be unique to prevent the communication from error.

2.Please contact Longer Company for communication protocol.

When the pump is idle, we recommend you to release the tubing
from pressure. This helps to protect the tubing from unnecessary
strain and prolongs its service life
Keep rollers clean and dry. This will prolong the service lives of
tubing and pump head.
The surface of drive and the pump head are not organic solvent
and aggressive liquids resistant. Please pay attention when using.

Note

If a trouble happens please contact us or our dealers.

The warranty period for this product is one year. If repair or
adjustment is necessary within the warranty period, the problem
will be corrected at no charge if it is not due to misuse or abuse on
your part, as determined by the manufacturer. Repair costs
outside the warranty period, or those resulting from product
misuse or abuse, may be invoiced to you.

,

Technical Specifications

Functions

Applicable Pump Head:
LEAD15-24, LEAD15-44, LEAD15-48, LEAD15-88
Flow Rate Adjustment:
Realize accurate flow rate adjustment function

Flow Rate Calibration:
Calibrate the setting flow rate to get accurate flow rate

Speed Control:
Realize accurate speed control through rotary coded switch

Display: 128

32 graphic LCD displays current running status

Prime: For fast filling and emptying

Direction Control: cw and ccw reversible

External Control: Start/stop control, direction control, speed

control (0 - 5V, 0 - 10V, 4 - 20mA optional)

Communication Function: RS485, Baud rate: 1200 bps

Memory Function: Storing the running parameter automatically

Cooling Mode: Heat-emitting fan

Specifications

Speed: 0.1 - 100.0 rpm, reversible

Speed Precision: 0.1 rpm

Adjusting Mode: Continuous rotary coded switch adjustment

Display Mode: 128 32 graphic LCD display. Display can switch

between flow rate and speed.

Applicable Power: AC 220V +/- 10%

50Hz+/-1Hz

AC 110V +/- 10%

60Hz+/-1Hz

Power Consumption:

50 W

Operating Condition: Temperature

0 to 40

Relative humidity < 80%

Drive Dimensions (L × W × H): 190 × 162 × 275 (mm)

Drive Weights: 3.5 Kg

IP Rating: IP 31

×

×

<

Содержание LEAD-2 Longer

Страница 1: ...g Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longer longerpump com Www longerpump com LEAD 2 PERISTALTIC PUMP OPERATING MANUAL ...

Страница 2: ...wn the power supply and pull out the plug when meet below circumstances Hold the plug not the power line when pulling out The fluids splash on the body of the pump You think the pump needs to be maintained or repaired Please shut down the power supply before install the external control equipments LEAD 2 OPERATING MANUAL Introduction Operation Panel Basic Operation Running Interface Menu Frame Wor...

Страница 3: ...wn the power supply and pull out the plug when meet below circumstances Hold the plug not the power line when pulling out The fluids splash on the body of the pump You think the pump needs to be maintained or repaired Please shut down the power supply before install the external control equipments LEAD 2 OPERATING MANUAL Introduction Operation Panel Basic Operation Running Interface Menu Frame Wor...

Страница 4: ... Control Interface P3 Power Socket Reference For detail information about external control function please see page 13 Basic Operation Start Stop Key Press the to start or stop the pump Direction Key Press the to change the rotation direction of the pump This key is valid when the pump is running Adjusting Knob Function 1 Adjust the flow rates or speed in running state Function 2 In menus selectio...

Страница 5: ... Control Interface P3 Power Socket Reference For detail information about external control function please see page 13 Basic Operation Start Stop Key Press the to start or stop the pump Direction Key Press the to change the rotation direction of the pump This key is valid when the pump is running Adjusting Knob Function 1 Adjust the flow rates or speed in running state Function 2 In menus selectio...

Страница 6: ...setting interface Note In speed Display the calibration function is forbidden In Running Interface press the it will display Adjusting Knob LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 3 Control mode 4 xit E 1 Calibration a Running direction b Current flow rate and speed c Tubing inner diameter ID0 5 means the inner diameter of the tubing is 0 5 mm d Start Stop s...

Страница 7: ...setting interface Note In speed Display the calibration function is forbidden In Running Interface press the it will display Adjusting Knob LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 3 Control mode 4 xit E 1 Calibration a Running direction b Current flow rate and speed c Tubing inner diameter ID0 5 means the inner diameter of the tubing is 0 5 mm d Start Stop s...

Страница 8: ... Knob Adjusting Knob Select Tubing ID 1 6 4 3 4 8 ID 0 5 INT 123 4mL min ID 0 5 INT 123 4mL min 5 6 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 1 Calibration 1 Calibration 2 ystem Setting S 2 ystem Setting S 4 xit E 4 xit E Act Vol 1234 μL Press to exit Test time 0 5min Cal Vol 1234 μL Press to begin Re calibrate Adjust Vol Act Vol 1234 μL Press to exit 1 Calibration a Tubing ID ...

Страница 9: ... Knob Adjusting Knob Select Tubing ID 1 6 4 3 4 8 ID 0 5 INT 123 4mL min ID 0 5 INT 123 4mL min 5 6 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 1 Calibration 1 Calibration 2 ystem Setting S 2 ystem Setting S 4 xit E 4 xit E Act Vol 1234 μL Press to exit Test time 0 5min Cal Vol 1234 μL Press to begin Re calibrate Adjust Vol Act Vol 1234 μL Press to exit 1 Calibration a Tubing ID ...

Страница 10: ...tem Setting interface press and then turn the to enter e Footswitch interface Press the to enter footswitch setting interface Turn the to select footswitch working mode Press the to save and return back to e Footswitch interface Trigger Press footswitch the pump starts running Press footswitch again the pump stops Gated The pump runs as long as the footswitch is pressed Adjusting Knob Adjusting Kn...

Страница 11: ...tem Setting interface press and then turn the to enter e Footswitch interface Press the to enter footswitch setting interface Turn the to select footswitch working mode Press the to save and return back to e Footswitch interface Trigger Press footswitch the pump starts running Press footswitch again the pump stops Gated The pump runs as long as the footswitch is pressed Adjusting Knob Adjusting Kn...

Страница 12: ...ly When the pump connects the footswitch the on operating panel is invalid no matter the remote control is on or off Start Stop Key ID 0 5 INT 123 4mL min Acceptable Pump Heads and Tubing Pump Head Tubing mm 16 25 17 LEAD15 48 LEAD15 88 LEAD15 24 LEAD15 44 2 4 Inner Diameter 1 0 Wall Thickness 16 25 17 9 10 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 1 Calibrati...

Страница 13: ...ly When the pump connects the footswitch the on operating panel is invalid no matter the remote control is on or off Start Stop Key ID 0 5 INT 123 4mL min Acceptable Pump Heads and Tubing Pump Head Tubing mm 16 25 17 LEAD15 48 LEAD15 88 LEAD15 24 LEAD15 44 2 4 Inner Diameter 1 0 Wall Thickness 16 25 17 9 10 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL 1 Calibration 2 ystem Setting S 1 Calibrati...

Страница 14: ...blow to the other side of the tubing If there is not air bubble in water the screws doesn t need to be adjusted If there is air bubble in water adjust the screws until the location where there is not air bubble in water 11 12 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL Pump Head Mounting The pump head is mounted on drive before the pump leaves the factory Follow below procedure for changing an...

Страница 15: ...blow to the other side of the tubing If there is not air bubble in water the screws doesn t need to be adjusted If there is air bubble in water adjust the screws until the location where there is not air bubble in water 11 12 LEAD 2 OPERATING MANUAL LEAD 2 OPERATING MANUAL Pump Head Mounting The pump head is mounted on drive before the pump leaves the factory Follow below procedure for changing an...

Страница 16: ...s below Please contact Longer Company for communication protocol Terminal Definition of Standard External Control Module Analog speed control input for external speed control in external control modules of 4 to 20 mA 0 to 5V and 0 to 10 V External start stop control input When connected to COM the pump runs when connected to high level 5 to 12 V the pump stops External cw ccw control input When co...

Страница 17: ...s below Please contact Longer Company for communication protocol Terminal Definition of Standard External Control Module Analog speed control input for external speed control in external control modules of 4 to 20 mA 0 to 5V and 0 to 10 V External start stop control input When connected to COM the pump runs when connected to high level 5 to 12 V the pump stops External cw ccw control input When co...

Страница 18: ... If repair or adjustment is necessary within the warranty period the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the manufacturer Repair costs outside the warranty period or those resulting from product misuse or abuse may be invoiced to you Technical Specifications Functions Applicable Pump Head LEAD15 24 LEAD15 44 LEAD15 48 LEAD15 88 F...

Страница 19: ... If repair or adjustment is necessary within the warranty period the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the manufacturer Repair costs outside the warranty period or those resulting from product misuse or abuse may be invoiced to you Technical Specifications Functions Applicable Pump Head LEAD15 24 LEAD15 44 LEAD15 48 LEAD15 88 F...

Страница 20: ...g Longer Precision Pump Co Ltd ADD Building A Chuangye Center Baoding National High Tech Industrial Development Zone Baoding Hebei China 071051 TEL 86 312 3110087 3138553 FAX 86 312 3168553 E mail longer longerpump com Www longerpump com LEAD 2 PERISTALTIC PUMP OPERATING MANUAL ...

Отзывы: