background image

Measurement   

I   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
5. 

Press the 

Enter

 key, the meter begins calibration. The Calibrating 

 

icon shows on the upper right corner of the screen. 

 

 

 

 
 

When the reading stabilizes, the meter will automatically show 

 

next calibration point. 

 

 

 

 
 

If necessary, press the 

 key to select a desired calibration point. 

 

 

 

 
6. 

Repeat steps 3 through 5 until the screen shows "Calibration is 

 

Completed", the meter returns to the measurement mode. 

 

 

To exit the calibration without saving calibrated values, press the 

ESC

 

key. 
 

 

 

Turbidity Measurement 

An accurate turbidity measurement depends on good measurement 
techniques. Factors such as clean sample vial, positioning of vial in 
the measurement chamber, covering the vial with the light shield lid, 
meter calibration, removal of bubbles, etc. Refer to the 

Measurement 

and Calibration Hints

 section for detailed instructions on page 3. 

 

Measuring the Low Turbidity Sample 

For the low turbidity samples (< 200 NTU), we recommend that using 
the same vial to perform the calibration and measurement. 
1.1  Rinse the vial with approximately 10 ml of sample. Cap and upend 
 

the vial several times. Discard the used sample and repeat the 

 

rinsing procedure two more times. 

1.2  Fill the vial with the sample. Cap the vial and wipe with a lint-
 

free cloth to remove waterdrops and fingerprints. Hold the vial 

 

cap and gently inverting vial several times, ensure that the sample 

 

is homogeneous and no air bubbles. 

 

 

 

 
1.3  Insert the vial into the measurement chamber. Align the triangle 
 

mark on the vial with the arrow on the meter. Close the light   

 

shield lid. 

 

 

 
1.4  Press the 

Meas

 key, the meter begins measurement, the

 

 

 

icon shows on the screen. When the reading stabilizes, the icon 

 

will automatically switch to

 

 

 

 
1.5  If necessary, press the 

Meas

 key to take a new measurement. 

 

 

During the measurement, press the   key, the reading will be locked. 
Press the key again to resume measurement. 

Align arrows 

Содержание TB200

Страница 1: ...TB200 Laboratory Turbidity Meter USER MANUAL Bante Instruments Inc ...

Страница 2: ...zin standard solutions 0 02 200 500 1000 NTU Two sample vials Lint free cloth 12V DC power adapter Meter Overview 1 Light shield lid 2 Measurement chamber 3 Display 4 Membrane keypad Connectors 1 Socket for power adapter 2 USB B interface to computer Keypad Key Function Switch the meter on or off Exit the calibration settings data logs and return to the measurement mode Start calibration Press and...

Страница 3: ...calibration Enable Disable default Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable Disable default Setting a Default Option 1 1 Press and hold the key to enter the setup menu 1 2 Press the key to select a menu item or an option 1 3 Press the Enter key to save To exit the setting without saving changes press the ESC key Setting the Password The password ...

Страница 4: ...ow on the meter Always close the light shield lid to prevent the measurement errors from ambient light Wash the sample vial with distilled water after measurement Any residues can cause inaccurate readings Turbidity Calibration The TB200 turbidity meter allows turbidity calibration up to 7 points with minimum of 2 points the default calibration points include the 0 02 10 200 500 1000 1500 and 2000...

Страница 5: ...n page 3 Measuring the Low Turbidity Sample For the low turbidity samples 200 NTU we recommend that using the same vial to perform the calibration and measurement 1 1 Rinsethe vial withapproximately10 mlof sample Cap andupend the vial several times Discard the used sample and repeat the rinsing procedure two more times 1 2 Fill the vial with the sample Cap the vial and wipe with a lint free cloth ...

Страница 6: ...hem decrease to room temperature 3 Weigh the filter disk and watch glass and record the mg value as B 4 Filter the 100 ml of sample through pre weighed filter disk Put the filter disk and watch glass in a 104 C 219 F drying oven for 1 hour 5 Remove the filter disk and watch glass and put in a desiccator Immediately cover the desiccator Wait for them decrease to room temperature 6 Weigh the filter ...

Страница 7: ...g the Data 1 Connect the USB cable to the meter and computer 2 Click the DAS icon the system automatically scans an available communication port and shows a message box Found a port on your computer 3 Click the OK the application starts 4 Click the Connect the screen shows Port is connected 5 Click the OK thenclick the Receive the storeddata will transfer to computer automatically If your computer...

Страница 8: ...100 ml volumetric flask and dilute to the mark 200 NTU Mix 10 ml of 4000 NTU standard in a 100 ml volumetric flask and dilute to the mark Mix 50ml of above standard in a 100 ml volumetric flask and dilute to the mark 10 NTU Mix 2 5ml of 200 NTU standard in a 100ml volumetric flask and dilute to the mark Or Mix 2ml of 500 NTU standard in a 100ml volumetric flask and dilute to the mark Optional Acce...

Страница 9: ...ter is one year from the date of shipment Above warranty does not cover the turbidity standards and glass vials Out of warranty products will be repaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Отзывы: