SALscan10/20 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
SALscan20 

 
Measurement Range 

Standard Solution Range 

Default 

200 to 2000 µS/cm 

700 to 1700 µS/cm 

1413 µS/cm 

2 to 20 mS/cm 

7 to 17 mS/cm 

12.88 mS/cm 

20 to 200 mS/cm 

70 to 170 mS/cm 

111.8 mS/cm 

 
Make sure that using the fresh standard solution during the calibration. 
DO NOT reuse the standard solution after calibration, contaminants 
in solution will affect the calibration and eventually the accuracy of 
the measurement. 
 

Single Point Calibration 

Ensure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the tester shows ----/CAL 1. 

 

 

 

 

1.2  Rinse the electrode with distilled water and place into the   
 

standard solution. Stir tester gently to remove air bubbles trapped

.

 

 

in the slot of the sensor. The tester will automatically recognize

.

 

 

the standard solution and show the calibration value. 

 

 

 

 

1.3  Press the 

Enter

 key, the default calibration value begins flashing. 

1.4  If necessary, press the 

Cal

 key to modify the calibration value, 

 

press the 

Enter

 key to confirm and move to the next digit. When 

 

the setting is completed, make sure that the displayed value   

 

matches the calibration standard. 

 

 

 

 

1.5  Press the 

Enter

 key, the tester begins the calibration. When the 

 

reading has stabilized, the display will show 

. Calibration 

 

is completed. 

 

 
 

 

 

 

Multipoint Calibration 

Ensure that you have selected 2 or 3 points calibration in the setup 
menu. 
2.1  Repeat steps 1.1 through 1.5 above. When the first calibration 
 

point is completed, the display will show ----/CAL 2, the tester 

 

prompts you to continue with second point calibration. 

2.2  Repeat steps 1.2 through 1.5 above until the display shows 

 

Calibration is completed. 

 

 

 

Performing the conductivity calibration will simultaneously   

 

calibrate the corresponding salinity value.   

 

To exit the calibration without saving calibrated values, press the 

 

Meas

 key. 

 

 

 

Temperature Calibration 

The SALscan series tester is installed with a built-in temperature sensor 
for automatic temperature compensation. During the measurement, 
if the measured temperature reading differs from that of an accurate 
thermometer, the tester needs to be calibrated. 
1. 

Press and hold the 

Cal

 key to enter the setup menu. 

 

 

 

 
2. 

Press the 

Enter

 key until the display shows

 

/

 

 

or

 

/

 

 

 

 

 

3. 

Press the 

Cal

 key, the tester enters the temperature calibration 

 

mode. 

 

Содержание SALscan10

Страница 1: ...SALscan10 20 Pocket Salinity Tester USER MANUAL Bante Instruments Inc ...

Страница 2: ...tion Confirm the calibration settings or displayed option Toggle between salinity and conductivity measurement modes Display Icon Description When the battery voltage falls below the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the...

Страница 3: ...hold the Cal key to enter the setup menu 2 If necessary press the Cal key again to select an option 3 Press the Enter key the tester saves the current option and moves to the next menu item 4 Repeat steps above until the tester returns to the measurement mode During the setting process if you do not need to calibrate the temperature press the Enter key to skip the or option To exit the setup menu ...

Страница 4: ...When the setting is completed make sure that the displayed value matches the calibration standard 1 5 Press the Enter key the tester begins the calibration When the reading has stabilized the display will show Calibration is completed Multipoint Calibration Ensure that you have selected 2 or 3 points calibration in the setup menu 2 1 Repeat steps 1 1 through 1 5 above When the first calibration po...

Страница 5: ...he sensor surface Wait for the measurement to stabilize and record the reading During the measurement DO NOTcompletely immerse thetester in water If the display shows indicating the measurement exceeds the range remove the tester from the sample immediately If the option is enabled in the setup menu the tester will automatically lock the measurement endpoint and show HOLD icon Press the Meas key t...

Страница 6: ...SALscan20 tester cell constant K 10 measuring range 0 1 to 200 mS cm ECCS 84 Standard solution 84 μS cm 480 ml ECCS 1413 Standard solution 1413 μS cm 480 ml ECCS 1288 Standard solution 12 88 mS cm 480 ml ECCS 50 Standard solution 50 mS cm 480 ml ECCS 1118 Standard solution 111 8 mS cm 480 ml Tester Specifications Model SALscan10 SALscan20 Conductivity Range 0 to 20 mS cm 0 to 200 mS cm Resolution ...

Страница 7: ...ester is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Отзывы: