background image

SALscan10/20   

I

    2 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Setup Menu 

The SALscan series tester contains 6 menu items in the setup menu, 
the following table describes the functions of each option. 
 

Menu Item 

Option and Description 

 

Calibration Points 

Set the number of calibration points. 

 

1 point (default) 

 

2 points 

 

3 points 

 

Measurement Unit 

Set the default temperature unit. 

 

Degrees Celsius (default) 

 

Degrees Fahrenheit 

 

Temperature Calibration 

Refer to page 3. 

 

Range: 0 to 100°C 

 

Range: 0 to 210°F 

 

Auto-Hold 

If enabled, the tester will automatically sense and 
lock the measurement endpoint. 

 

Enable 

 

Disable (default) 

 

Auto-Power Off 

If enabled, the tester will automatically switch off 
if no key is pressed within 8 minutes. 

 

Enable (default) 

 

Disable 

 

Factory Reset 

If enabled, all of the calibration values and current   
settings will be deleted or reset to the factory   
defaults, the tester must be recalibrated. 

 

Enable 

 

Disable (default)   

 

Setting the Default Option 

1. 

In the measurement mode, press and hold the 

Cal

 key to enter 

 

the setup menu. 

 

 

 

 
2. 

If necessary, press the 

Cal

 key again to select an option. 

 

 

 

 
3. 

Press the 

Enter

 key, the tester saves the current option and moves 

 

to the next menu item. 

 

 

 

 
4. 

Repeat steps above until the tester returns to the measurement 

 

mode. 

 

 

 

During the setting process, if you do not need to calibrate the 

 

temperature, press the 

Enter

 key to skip the

 

/

 

 

or

 

/

 

 

 

option. 

 

To exit the setup menu, press the 

Meas

 key. 

 

 

 

Conductivity Calibration 

The SALscan series tester allows up to 3 points calibration. For better 
accuracy, we recommend that you perform 3 points calibration or   
select a standard solution closest to the sample conductivity you are 
measuring. The tester will automatically detect the calibration standard 
and prompt the user to perform the calibration. When the calibration 
is completed, all new calibration values will override existing data.The 
following table shows acceptable standard solutions for each tester. 
 
SALscan10 

 
Measurement Range 

Standard Solution Range 

Default 

0 to 200 µS/cm 

70 to 170 µS/cm 

84 µS/cm 

200 to 2000 µS/cm 

700 to 1700 µS/cm 

1413 µS/cm 

2 to 20 mS/cm 

7 to 17 mS/cm 

12.88 mS/cm 

 

Option 

Menu item 

Содержание SALscan10

Страница 1: ...SALscan10 20 Pocket Salinity Tester USER MANUAL Bante Instruments Inc ...

Страница 2: ...tion Confirm the calibration settings or displayed option Toggle between salinity and conductivity measurement modes Display Icon Description When the battery voltage falls below the minimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the...

Страница 3: ...hold the Cal key to enter the setup menu 2 If necessary press the Cal key again to select an option 3 Press the Enter key the tester saves the current option and moves to the next menu item 4 Repeat steps above until the tester returns to the measurement mode During the setting process if you do not need to calibrate the temperature press the Enter key to skip the or option To exit the setup menu ...

Страница 4: ...When the setting is completed make sure that the displayed value matches the calibration standard 1 5 Press the Enter key the tester begins the calibration When the reading has stabilized the display will show Calibration is completed Multipoint Calibration Ensure that you have selected 2 or 3 points calibration in the setup menu 2 1 Repeat steps 1 1 through 1 5 above When the first calibration po...

Страница 5: ...he sensor surface Wait for the measurement to stabilize and record the reading During the measurement DO NOTcompletely immerse thetester in water If the display shows indicating the measurement exceeds the range remove the tester from the sample immediately If the option is enabled in the setup menu the tester will automatically lock the measurement endpoint and show HOLD icon Press the Meas key t...

Страница 6: ...SALscan20 tester cell constant K 10 measuring range 0 1 to 200 mS cm ECCS 84 Standard solution 84 μS cm 480 ml ECCS 1413 Standard solution 1413 μS cm 480 ml ECCS 1288 Standard solution 12 88 mS cm 480 ml ECCS 50 Standard solution 50 mS cm 480 ml ECCS 1118 Standard solution 111 8 mS cm 480 ml Tester Specifications Model SALscan10 SALscan20 Conductivity Range 0 to 20 mS cm 0 to 200 mS cm Resolution ...

Страница 7: ...ester is one year from the date of shipment Above warranty does not cover the electrode and standard solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...

Отзывы: