pH Measurement

Rinse

 

the

 

electrode

 

thoroughly

 

with

 

distilled

 

water,

 

 

 

   

ait for the

to stabilize, record

    

 

    

 

Electrode Replacement

When the tester fails to calibrate or gives fluctuating readings for calibration 
standards, you need to replace the electrode module.
1. Twist the electrode collar counter clockwise, 

 

pull the old electrode module

    away from the tester.

2. Align the slot on the new electrode

 

module,

 

gently

 

push

 

the

 

module

 

into

 

the

    tester.

3. Twist the electrode

 

collar

 

clockwise until it is tight. 

1. 

press MEAS key to turn

    on the tester.
2. Dip the electrode into the sample solution, stir the tester gently.
3. W  

 

 reading   

 the measured value on the display.

Hold Function

1. During 

 

the measurement, press HOLD key, 

 

, “HOLD” indicator appears on the display.

2. Press the key again to release 

 

the measured 

 

reading, 

 

you can continue to

    take measurement.

Electrode Care and Maintenance

  For best results, please always keep the sensor’s glass bulb wet.

  

Ensure

 

that

 

the

 

sensor

 

is

 

thoroughly

 

washed

 

with

 

distilled

 

water

 

after

 

each

use.

  If 

 

you do not use 

 

the tester for long periods, please store the sensor with

electrode storage solution. DO NOT use de-ionized or distilled water.

 

 

Installation

 

is

 

completed.

the tester immediately freezes

    current displayed value

Troubleshooting

 Electrode dried out.

   

Soak the

 

electrode 

 

in

 

tap

 

water

 

for

 

10

 

minutes. 

 Measured value is out of range.

   Check the electrode whether clogged, dirty or broken.

module

 Incorrect buffer used or contaminated buffer solution.

   Use fresh pH buffer solutions for calibration.

 Electrode is broken.

   Replace the electrode 

.

module

LCD DISPLAY

CAUSE

Optional Electrode Modules

ORDER CODE

DESCRIPTION

E-PHSCAN-S

Standard pH Electrode
Used for routine pH measurements,
E.g. drinking water, rainwater, etc.

E-PHSCAN-F

Flat Surface pH Electrode
Used for measuring the skin, fruits, 
meats etc.

E-PHSCAN-L

Long-Reach pH Electrode
Used for measuring the samples in 
the test tube.

Specifications

Model

pH Range

Accuracy

Calibration Points

Calibration Solutions

Operating Temperature

PHscan10

0.0~14.0pH

±0.1pH

2 points

USA (pH4.01, 7.00, 10.01)

o

o

0~60 C, 32~140 F 

Power Requirement

Dimensions

Weight

3 1.5V “

Batteries

×

G13A” 

185(L) 40(Dia.)mm

×

100 grams